- Phone
-
Address
No. 4 Yuantou Road, Licang District, Qingdao City, Shandong Province
Qingdao Shangde Environmental Protection Technology Co., Ltd
No. 4 Yuantou Road, Licang District, Qingdao City, Shandong Province
SN-200Y-F1type
portablemulti-parameterwater qualitydetector
(Optional:CODAmmonia nitrogen, total phosphorus, total nitrogen)
1Instrument Description
SN-200Y-F1 portable COD multi parameter water quality analyzerAdopting a touch sensitive design, a fully Chinese operating interface, and guided system operation, it can perform digestion pretreatment and measurement of water quality indicators after pretreatment is completed.
SN-200Y-F1 portable COD multi parameter water quality analyzerThe whole machine is designed to be portable, using a dual-purpose tube for digestion and colorimetry360The ° rotation colorimetric technology has high detection safety and accuracy, and can be used for outdoor on-site detection of common water qualityCODAmmonia nitrogen, total phosphorus, total nitrogenPollutants can also be used for daily water quality testing in the laboratory.
The instrument is widely used in environmental monitoring, scientific research water quality testing, production emission monitoring and testing, and other fields.
IIMain functional features
(1)DoubleColor touch screen design, friendly human-machine dialogue interface, easy to operate, direct reading of concentration;
(2)Integrated design for digestion and measurement, zone control, capable of simultaneous digestion and measurement of water quality indicators;
(3)The portable box design of the measuring instrument has a high level of protection, is waterproof and dustproof, and is convenient for carrying out measurements of water quality indicators;
(4)Schematic diagram of panel configuration operation process, convenient for users to perform quick measurement operations;
(5)Equipped with a large capacity lithium battery, it can complete multiple digestion and measurement tasks, meeting the needs of field measurements;
(6)Using a dual-purpose digestion colorimetric tube makes the water sample determination process simple, fast, and safe,360The ° rotation colorimetric technology makes the measurement results more accurate;
(7)Large capacity data storage with power-off protection design, 以防Instrument damage and data loss;
(8)1point to7Point correction mode, automatically calculates slope, intercept, and correlation coefficient, with high measurement accuracy;
(9)Can connect to a computer to export measurement data, comes with a printer, can print measurement data and historical records;
(10)Low reagent usage, low operating costs, and strong anti-interference ability;
(11)The concave digestion zone design effectively protects the safety of operators;
(12)Pre stored standard measurement curves can be easily retrieved for measurement, and self built measurement curves can also be created. Multiple indicator digestion programs can be pre stored, and digestion can be started with one click.
IIIperformance parameters
Measurement indicators:optionalCODAmmonia nitrogen, total phosphorus, total nitrogen;
Display mode: dual color LCD touch screen, Chinese operation interface;
Colorimetric method: Colorimetric tubecolorimetry;
Number of curves: adjustable100Measurement curve parameters;
Pre stored digestion program:CODTotal phosphorus and total nitrogen digestion program, with two custom programs;
Simultaneous digestion sample size: ≤6Support;
Temperature indication error: ≤±2℃;
Record storage: storable10000The test results show that the data will not be lost when the power is cut off;
hit Printing: Comes with a thermal printer;
Data communication:USBConfigure dedicated computer software to upload measurement results to the computer for user statistical analysis;
Environmental temperature:(5~40)℃;
Environmental humidity: Relative humidity ≤85%(No condensation);
Charging power supply:AC220V±15% / 50Hz;
Power supplyDC16.8VBuilt in rechargeable lithium battery;
Overall dimensions:420mm×350mm×180mm;
Instrument weight: approximately7Kg.
indicator name |
measurement range |
measurement error |
COD |
0~15000 mg/L |
≤5% |
ammonia nitrogen |
0~50mg/L |
≤5% |
Total Phosphorus |
0~20mg/L |
≤5% |
Total Nitrogen |
0~100mg/L |
≤5% |