Repair of Suzhou Mal roughness tester PS1Suitable for production sites, it can measure the surface roughness of various machined parts. According to the selected measurement conditions, the corresponding parameters of the computer can be displayed clearly on the LCD monitor, and all measurement results and graphics can be output on the printer. The surface roughness meter can also communicate with a PC.



Repair of Suzhou Mal roughness tester PS1andMaintenance:
(1) Strictly avoid collisions, severe vibrations, heavy dust, moisture, oil stains, strong magnetism, etc.
(2) Please store the sensor in the box promptly after use.
(3) When the battery voltage is insufficient, it should be charged in a timely manner. It is allowed to insert a power adapter while working, but testing samples with low Ra values will affect the testing accuracy. If the voltage is still insufficient after charging for several hours or if it is found to be insufficient after a short period of use after being fully charged, the battery needs to be replaced.
(4) Due to the fact that sensors are highly precise components, careless disassembly and assembly can result in damage. Therefore, it is recommended to use them in a centralized manner during measurement and minimize the number of disassembly and assembly to avoid affecting the maintenance and repair of the roughness meter.
(5) When the measurement error exceeds the range of ± 10% and the user confirms that the error is not caused by human factors, the software calibration function specially designed for this machine can be used. The calibration value is a percentage that represents the percentage increase in the measured results after calibration relative to those without calibration.
(6) When abnormal phenomena occur, users can try to repair them without disassembling the device. If it cannot be repaired, please submit it to the repair department. Users should not disassemble the roughness meter themselves, so that the company can perform maintenance and repair checks on it.