I provided the FN-ATA-16 RF probe for the FN-DGZ-1120 electromagnetic radiation comprehensive tester,
Mainly used in the following areas: substations, transmission lines, and power equipment; Railway, subway, and related rail transit; Base stations, navigation stations, and related communication equipment; Long, medium, and short waves, broadcasting and television signals; Various household appliances; Industrial, medical, and experimental equipment
Electromagnetic Environment Control Limits (GB8702-2014)
Guidelines for Radiation Environmental Protection Management - Electromagnetic Radiation Monitoring Instruments and MethodsHJT 10.2-1996)
Methods and Standards for Environmental Impact Assessment of Electromagnetic Radiation (HJ/T10.3-1996)
Electromagnetic Environment Monitoring Method for Communication Transmission and Transformation Engineering(HJ 6812013)
Guidelines for Limiting Exposure to Time Varying Electric, Magnetic, and Electromagnetic Fields (300GHz)below)》(ICNIRP1998Guidelines)
Monitoring Method for Electromagnetic Radiation Environment of Mobile Communication Base Stations(HJ972-2018)
The performance is as follows:
FN-ATA-16Probe(100kHz-6GHz)
frequency range |
100kHz-6GHz |
range |
0.2V/m~300V/m(1.06×10-4W/m2~238W/m2) |
overload protection |
600V/m |
Frequency error |
<10 of the tested frequency-3order of magnitude |
frequency response |
900MHz~3GHz,≤±1.5dB;<900MHz, or>3GHz,≤±2dB |
dynamic range |
>60dB |
linearity |
≤±1.5dB |
isotropic |
<1.5dB |
Antenna diameter |
95mm |