- Phone
-
Address
9F, Zone C, Jiujiang Shuixin Hongchangping Zhigu, Changping Town, Dongguan City
Dongguan Qiya Testing Equipment Co., Ltd
9F, Zone C, Jiujiang Shuixin Hongchangping Zhigu, Changping Town, Dongguan City
Technical Scheme for Thermal Aging Test Chamber
*Control methods and characteristics:
Balanced temperature and humidity control system(BTHC), in order toP.I.D.Method controlSSRMake the heating capacity of the system equal to the moisture and heat loss, so it can be used stably for a long time.
Thermal aging test chamberUsed for conducting thermal aging tests on plastic materials (rubber, plastic), electrical insulation, and other materials in scientific research institutions and factories. conform toGB\\T 3512-2014.ISO188; 2011Waiting for national standards.
1、 Performance:At room temperature of 20 ℃, when unloaded
1.1Temperature range:RT+10℃~300℃
1.2Temperature fluctuation:±0.5℃
1.3 Temperature uniformity:±2℃
1. 4 Air exchange rate: 2-100 times/H
1.5 Timing devices0~999hour
1.6 Wind speed: adjustable from 0m/s to 2.0m/s
1.7test stand:rotation speed10RPM/minTest rack
2、 Structure and Material:
machine type |
inside ministry ruler inch |
outside ministry ruler inch |
QYD-150L |
500x600x500(WxHxD)MM |
850x1600x900(WxHxD)MM |
2.1Inner box material:stainless steel plate
2.2outer box material:SPCC board with electrostatic baking paint
2.3Insulation material:glass wool
2.4Supply air circulation system:
2.4.1 60WMotor,1only
2.4.2Stainless steel extended axis
2.4.3Multi wing fan blade(SIRCCO FAN)
2.4.4Adjustable louvers ensure even temperature distribution inside the test chamber
2.5box door:Single panel door, left opening, handle on the right-hand side
2.5.1Flat embedded handle
2.6ventilation system:
2.6.1 25WMulti wing snail fan
2.6.2A code tuner
2.6.3One adjustable air door
3、 Control system:
3.1 Temperature Controller:Original digital display button controller
3.2 power meter:Digital display button power meter
3.2 heating system:
3.2.1heater:Nickel chromium alloy finned heat dissipation tubeUHigh efficiency heater with automatic calculation and high precision
degreePID+SSRControl (calculation cycle:1~4S can be set up)Make its heat loss equal to the heating amount
Comparable, therefore more than traditionalON/OFFControl power saving approximately30%
3.3 Temperature measuring body: DINspecificationlevel ψ4.8mm SUS #304Stainless steelPT 100Ω 1branch
3.4 safety device:
3.4.1Zero crossing gate fluid power controller
3.4.2Air burning prevention protector
3.4.3Underphase reverse phase protection switch
3.4.4 Line fuses and fully sheathed terminals
3.4.5 buzzer
4、 Standard configuration:
4.1 power cord4.0 mm²x5Rui, low impedance rubber cable2.5mlong1item
4.2 One set of controller user manual and one set of operation manual
5、 Surrounding environment:
5.1Allowable temperature range for use0~35℃
5.2Performance guarantee scope:5~35℃
5.3 Relative humidity: not greater than85%
5.4Atmospheric pressure:86~106Kpa
5.5There is no strong vibration around
5.6No direct sunlight or other heat sources
6、 Use power supply:AC 3 µ 4W380V 50HZ; Power: 6.0KW