Performance characteristics:
The sterilization chamber adoptsS30408Stainless steel, the jacket is made of specialized steel plates for boiler pressure vessels, and the main body is manufactured using advanced welding processes and testing methods to ensure reliable quality and durability of the sterilizer body;
The entire sterilization process is automatically controlled by a microcomputer, with a high degree of automation. The color touch screen can dynamically display parameters such as working temperature, pressure, time, and sterilization stage, making the operation more intuitive and convenient;
The equipment is equipped with multiple sterilization programs (fabrics, instruments, liquids, etc.), and working parameters can also be set according to the actual situation of the user;
Simultaneously equipped with vacuum detectionBDTest program, vacuum leakage rate less than0.13Kpa/min;
The sealed door adopts a safety interlock device that meets regulatory requirements:
(1)When the door is not closed and locked, the sterilization program cannot be started;
(2)When there is pressure in the interior, the sealing door cannot be opened;
The sealing ring adopts a compressed air sealing structure design, which has the characteristics of good sealing effect and long service life;
PIDThe temperature control function can automatically adjust the heating control parameters by selecting, preventing temperature overshoot from occurring.
Note: The sterilizer for self-produced steam is equipped with a steam generator and a water pump, which have the following functions:
(1)The automatic water inlet and replenishment function of the water pump ensures that the water level in the steam generator is controlled within a reasonable range;
(2)When the water level in the steam generator reaches the lowest value, it can cut off the power supply of the electric heating tube and give an alarm prompt;
(3)The pressure inside the steam generator is controlled by a pressure controller, and there is a safety valve for protection after overpressure.
Main technical parameters
|
Interior design pressure |
-0.1~0.26MPa |
Interlayer design pressure |
0.26 MPa |
|
Rated working pressure |
0.242MPa |
Steam source pressure |
0.30MPa~0.50MPa |
|
Compressed air source pressure |
0.50MPa~0.70MPa |
Water source pressure |
0.15MPa~0.30MPa |
|
Design maximum operating temperature |
140℃ |
Rated operating temperature |
138℃ |
|
Pulse frequency range |
0~99time |
Pulse amplitude |
at the time of0.065MPa~0.09MPa |
|
Drying time range |
0~9999second |
Air discharge volume |
>99%(Pulse frequency)3More than once) |
|
Setting pressure of safety valve |
0.25MPa |
Return pressure |
0.22MPa |