Programmable cold and hot alternating test chamberProduct Features
The temperature control adopts a fully imported touch button instrument, with simple operation and setting.
After inputting the data and experimental conditions, the controller has a locking function to prevent human touch from changing the temperature value.
Equipped with P.I.D automatic calculation function, it can immediately correct temperature change conditions, making temperature control more precise and stable.
The main component of the electrical control adopts the Korean TEMI300 instrument to better control the temperature
Programmable cold and hot alternating test chamberParameter Specifications:
|
The high and low temperature alternating test chamber prohibits the testing and storage of flammable, explosive, and volatile substance samples, corrosive substance samples, biological samples, and strong electromagnetic emission source samples. |
|||
|
Common models |
DEJG-010A type |
||
|
Studio size (D × W × H) |
1000×1000×1000 mm |
||
|
Dimensions (D × W × H) |
1480×1480×1850mm |
||
|
total power |
About 7.5 KW |
||
|
Main performance parameters
|
temperature range |
A:-20℃~150℃ B:-40℃~150℃ C:-60℃~150℃ D:-80℃~150℃ |
|
|
Temperature deviation |
≤±2.0℃ |
||
|
temperature fluctuation |
≤±0.5℃ |
||
|
Heating and cooling rate |
heating up time |
2-3 ℃/min (when the equipment is unloaded) |
|
|
Cooling time |
0.7~1 ℃/min (when the equipment is unloaded) |
||
|
volume |
rise |
Xiamen Deyi CompanyHigh and low temperature alternating machineThere are also capacities of 80L, 100L, 150L, 250L, 408L, 500L, 600L, 800L, and 1000L available for purchase! |
|
|
Performance characteristics |
|
||
|
monitor |
5.7 "320 × 240 dot matrix color LCD display. (TEMI880, South Korea) |
||
|
Operation mode |
Fixed value or program |
||
|
Setting method |
Chinese menu, touch screen input |
||
|
Program capacity |
Can store 120 sets and 1200 segments of test procedures |
||
|
program length |
Each program has an additional 99 segments |
||
|
Number of cycles |
Each program segment can loop 999 times |
||
|
Setting Range |
Temperature: 0 ℃~+150 ℃ |
||
|
Time: 0-999 hours and 59 minutes |
|||
|
resolution |
Temperature: 0.1 ℃ |
||
|
Time: 0.1min |
|||
|
input |
PT100 platinum resistor |
||
|
communication |
Equipped with an RS232/485 interface |
||
|
Auxiliary functions |
Alarm display function (indicating the cause of the fault), power outage protection function, upper and lower limit temperature and humidity alarm function, timing function (automatic start and stop), self diagnostic function. |
||
|
Measurement data collection |
PT100 platinum resistor |
||
|
Cabinet material |
Inner box material |
Stainless steel plate |
|
|
Outer box material |
Stainless steel plate |
||
|
thermal insulating material |
Rigid polyurethane foam+glass fiber |
||
|
door |
Single door, equipped with observation window, lighting, left hinge and right handle |
||
|
refrigeration systemProduct advantages |
refrigeration compressor |
Original French imported "Taikang fully enclosed piston compressor" |
|
|
Cooling method |
Compression adopts a cascade refrigeration system for cooling |
||
|
Cooling method |
forced air cooling |
||
|
evaporative condenser |
Brazed plate heat exchanger |
||
|
condenser |
air-cooled condenser |
||
|
Control mode of refrigeration machine |
The PLC (Programmable Logic Controller) of the control system automatically selects and adjusts the operating conditions of the refrigeration machine according to the test conditions Condensation pressure regulating valve Evaporator Pressure Regulator Compressor return air cooling circuit Energy regulation circuit |
||
|
refrigerant |
Using green and environmentally friendly refrigerants R404A and R23 as the refrigeration medium for this refrigeration system |
||
|
Main refrigeration components |
Main refrigeration accessories: solenoid valve (SANHUA, Japan), filter (Danfoss, Denmark), pressure controller (Danfoss, Denmark), oil separator (Aiko, USA), plate heat exchanger (Shuripe, Sweden), etc. are all imported brand products. |
||
|
**Protective device |
Leakage protector |
Compressor overpressure protector |
|
|
Over temperature protector inside the box |
Compressor overheat protector |
||
|
Fan overheat protector |
Compressor overcurrent protector |
||
|
phase sequence relay |
|
||
|
Air conditioning channel A |
Centrifugal fans: 2 units |
||
|
Standard configuration of test chamberProduct Information
|
Observation window: 1 transparent electric heating film insulated tempered glass (located on the door) The visual range is approximately 280mm × 350mm, including the range of lighting obstruction. |
||
|
Lead hole: 1 hole with a diameter of 50mm (located on the left side of the box) |
|||
|
Indoor lighting fixture on the door: 1 36W/220V |
|||
|
Mobile wheels: 4 |
|||
|
Test hole: 1 50mm hole on the left side of the body, with 1 stainless steel hole cover and 1 silicone plug. |
|||
|
Power cord 5m |
|||
|
Sample rack: 2 layers of stainless steel sample rack, load-bearing capacity (evenly distributed): 25kg/layer (The cumulative total load capacity of the samples in the box shall not exceed 100kg) |
|||
|
Conditions of Use |
power supply |
○ Voltage allowable fluctuation range: AC220V ± 10% ○ Allowable frequency fluctuation range: 50Hz ± 0.5 Hz The grounding resistance of the protective grounding wire is less than 4 Ω; TN-S power supply or TT power supply |
|
|
Usage environment |
Temperature: 5 ℃~35 ℃, relative humidity: ≤ 85% RH |
||
|
site |
○ The ground is flat, well ventilated, and free of flammable, explosive, corrosive gases and dust There is no strong electromagnetic radiation source nearby ○ Leave appropriate maintenance space around the equipment |
||
|
Meet standards |
GB/T 2423.1-2008 Test A: Low temperature test method |
||
|
|
GB/T 2423.2-2008 Test B: High temperature test method |
||
|
|
GJB 150.3A-2009 High Temperature Test |
||
|
|
GJB 150.4A-2009 Low Temperature Test |
||
1. The inspection and adjustment of the wet bulb water level in the high and low temperature alternating humidity test chamber should not be too high in the water tank, causing water to overflow from the tank or too low, resulting in abnormal water absorption of the wet bulb test cloth and affecting the accuracy of the wet bulb. The water level should be kept at about six tenths full. The adjustment of the water level in the water tank can adjust the height of the water tank.
2. The replacement of the wet bulb test cloth in the high and low temperature alternating humidity test chamber is necessary when the surface of the test cloth is not clean or hardened, or before continuing to perform temperature and humidity bulb control after completing temperature control. The test cloth should be replaced about every three months. When replacing, the temperature sensor (Sen sor) should be wiped with a clean cloth. When replacing the new test cloth, it should be cleaned first.
3. The cleaning of dust on the condenser of the high and low temperature alternating humidity test chamber should be carried out regularly every month. The condenser should be cleaned with a vacuum cleaner to remove the dust attached to the condenser heat dissipation mesh or sprayed with high-pressure air to remove the dust.
4. During the operation of the high and low temperature cyclic testing machine, the setting of the over temperature protection in the high and low temperature alternating humidity test chamber should be increased by 20 ℃~30 ℃. When the temperature inside the test chamber rises to the set point for over temperature protection, the power supply to the heater stops, and the 'OVERHEAT' over temperature warning light lights up but the fan still runs. If it runs for a long time without supervision, please make sure to check the over temperature protector before operation to ensure that it is properly set (the wet bulb over temperature protector is set to 120 ℃).
High and low temperature alternating machineCleaning and maintenance inside and outside the body
1) The high and low temperature alternating humidity test chamber should be cleaned of internal impurities before operation.
2) The distribution room should be cleaned at least once a year, and a vacuum cleaner can be used to remove the indoor dust during cleaning
3) The exterior of the box should also be cleaned at least once a year, by wiping it with soapy water before cleaning.
High and low temperature alternating machineInspection and maintenance of humidifiers
The water storage in the humidifier should be replaced once a month to ensure clean water quality, and the humidification water tray should be cleaned once a month to ensure smooth water flow.

Xiamen Deyi Company's main products: high and low temperature alternating humidity test chamberHigh and low temperature alternating machineHigh and low temperature alternating test chamber, high and low temperature test chamber, high and low temperature impact test chamberHumidity ChamberStep into a constant temperature and humidity chamber, a low-temperature constant temperature and humidity chamber, a low-temperature low humidity chamber, a low-temperature constant temperature chamber, a ventilated aging chamber, a ventilated thermal aging chamber, a xenon lamp aging test chamber, an ultraviolet weather resistance test chamber, an ozone aging test chamber, a rain test chamber, a swinging tube rain test device, a strong water flushing test device, an anti immersion test device, a sand and dust test chamber, a salt spray test chamber, a yellowing resistance test chamber, an electromagnetic vibration test bench, a simulated automobile transportation vibration test bench, a single arm drop test machine, a double wing drop test machine, a zero drop test machine, a mobile phone drop test machineProgrammable cold and hot alternating test chamberEnvironmental testing equipment including carbon fiber material tensile test bench, single column micro controlled carbon fiber material tensile test bench, single column LCD display carbon fiber material tensile test bench, cardboard box strength testing machine, room temperature drying oven, vacuum drying oven, electric hot air drying oven, precision drying oven, glassware drying oven, etc.