Welcome Customer !

Membership

Help

Xiamen Deyi Equipment Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Products

Price of programmable temperature and humidity control test chamber

NegotiableUpdate on 06/30
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

Brief introduction: The programmable temperature and humidity control box is particularly suitable for fiber optics LCD、 Crystals, inductors PCB、 High temperature resistance, low temperature resistance, and moisture resistance cycling tests for products such as batteries, computers, and mobile phones.

Product Details

Details introduction:

Programmable temperature and humidity control experimental chamberProduct Introduction:This equipment is mainly designed for adaptability testing of electrical and electronic products, their raw components, and other materials during storage, transportation, and use in high and low temperature, humid and hot environments. This testing equipment is mainly used to conduct environmental simulation tests on the physical and other related characteristics of products under high and low temperature conditions in accordance with national standards or user defined requirements. After testing, the performance of the product is judged through inspection to determine whether it can still meet the predetermined requirements, for product design, improvement, identification, and factory inspection purposes


Programmable temperature and humidity control experimental chamberapplication area:


1. Simulate various climatic conditions and apply them to stability testing of samples such as drugs and food.

In the field of life sciences, it is applied to microbial cultivation and sample storage.

3. Aerospace products, information electronic instrumentsInstruments, materials, electricians, electronic products, and various electronic components are tested for their performance indicators in high and low temperature or humid and hot environments. Application fields: Constant temperature and humidity test chambers, also known as constant temperature and humidity testing machines, constant temperature and humidity test chambers, programmable humid and hot alternating test chambers, constant temperature machines or constant temperature and humidity chambers, are equipment used to test the performance of materials in various environments and test the heat resistance, cold resistance, dry resistance, and moisture resistance of various materials.


Programmable temperature and humidity control experimental box product album


Programmable temperature and humidity control experimental chamberParameter specifications

Common models

Studio Size(W*H*D)mm

External dimensions(W*H*D)mm

DEJS-100D

450×400×500

600×1400×1300

DEJS-150D

500×600×500

700×1500×1300

DEJS-225D

500×750×600

800×1750×160

DEJS-408D

600×850×800

800×850×800

DEJS-800D

800×1000×1000

1200×1900×1650

DEJS-010D

1000×1000×1000

1200×1900×1850

Performance characteristics

Test environment conditions

test method

The ambient temperature is+25℃, relative humidity ≤85%Under the condition of no sample in the test chamber

GB/T 5170.2-2008Temperature testing equipment;GB/T 5170.5-2008Damp heat testing equipment.

temperature range

-20℃→+150℃

temperature fluctuation

≤±0.5℃

Temperature deviation

≤±2℃

Humidity range

20%~98%R.H.

Humidity fluctuation

≤±2%R.H.

humidity uniformity

≤±3%

heating up time

-40℃→+150℃ make an appointment45 min Nonlinear no-load

Cooling time

+20℃→ -40℃ make an appointment60 min Nonlinear no-load

Temperature and humidity control range diagram

Meet the testing standards

GB/T11158-2008Technical conditions for high temperature test chamber;

GB/T10589-2008Technical conditions for low-temperature test chamber;

GB/T10592-2008Technical conditions for high and low temperature test chambers

GB/T10586-2006Technical conditions for wet heat test chamber;

GB/T2423.1-2008testALow temperature test method

GB/T2423.2-2008testBHigh temperature test method;

GB/T2423.3-2006Damp heat test chamber test method;

GB/T2423.4-2008Alternating damp heat test method;

GB/T2423.22-2002Temperature change test method;

IEC60068-2-1:2007 Low temperature test chamber test method;

IEC60068-2-2:2007 High temperature test chamber test method;

GJB150.3High temperature test; GJB150.4Low temperature test;GJB150.5Damp heat test;

IEC68-2-1testA: Cold; IEC68-2-2testB: Dry heat;

structure characteristics

Thermal insulation enclosure structure

Outer wall material: cold-rolled steel plate with high-quality baked paint.

Inner wall material:304stainless steel .

Box insulation material: rigid polyurethane foam+glass fibre.

Door insulation material: rigid polyurethane foam+glass fibre.

Air conditioning channel

Centrifugal fan: 1individual.

Heater, humidifier, evaporator, water supply and drainage outlet, temperature sensor.

Standard configuration of test chamber

1Group large window250x330x40mmthree

Vacuum layer;

Flat embedded handle;door hinge:SUS #304

;1Lamp inside the box(

A Philips observation light);stainless steelSUS #3042Grid style steel plate storage rack

Piece (adjustable up and down, detachable);Stainless steel adjustable spacing40mm2Storage rack track

Group;Lead hole: φ50mm 1

One.

Box dooropen-door,Drive from right to left,

Place your hand on the right-hand side;Equipped with observation windowW250Xh330

Mm, lighting fixtures;window frame/

Door frame anti condensation electric heating device.

control panel

Controller display screen, over temperature protection setter

machine room

Refrigeration unit, water tray, drainage hole

Distribution control cabinet

Distribution board, exhaust fan

heater

Heating system: finned heat dissipation tube shaped stainless steel electric heater heating air circulation method

humidifier

International advanced shallow trough boiler humidification system.

Power cord hole and drainage hole

Located on the back of the box

refrigeration system 1

way of working.

mechanical refrigeration 2

refrigeration compressor.

3Imported French Taikang low-temperature and high-efficiency fully enclosed compressor

Evaporator.

4Fin type multi-stage automatic load capacity adjustment

Condenser.

5Air-cooled separation type

Expansion system: Capacity controlled refrigeration system.

6. (Capillary tube)

evaporative condenser.

7. Stainless steel brazed plate heat exchanger

Control mode of refrigeration machine.

The controller of the control system automatically adjusts the operating conditions of the refrigeration machine according to the test conditions.

Evaporator Pressure Regulator

8. Compressor return air cooling circuit

refrigerantDupontR404a0Environmentally friendly refrigerant (with an ozone depletion index of)

)

1electrical control system

Independently developed control system.

2Human computer friendly interactive interface, intuitive temperature and humidity display

Temperature and humidity controllerAdopting imported microcomputer color LCD display touch screen direct button type,Wide perspective of Chinese and English expressions,High contrast function with temperature and humidity programmable controller,250utensil12500Group ProgramSegmental memory,each paragraph520Hour59Min,Can be arbitrarily divided and set,And attach multiple groupsPID.

3control function

Controller specificationsaccuracy:0.1Temperature ±℃+1digitHumidity ±1%R.H+1digit

;resolution:0.01Temperature ±℃,Humidity ±0.1%R.H

;Temperature slope:0.19.9~

Can be set;

Equipped with upper and lower limit standby and alarm functions;Temperature and humidity input signal dry wet bulbPT100x2

9;groupP.I.DControl parameter setting, P.I.D

Automatic calculation;.

4Dry wet bulb automatic calibration

Screen display functionColor Painting Face to Face Discussion Style,No need for button input,

Screen direct touch options;Temperature and humidity setting(SV)Compared to reality(PV

)Value displayed directly;Can display the current program execution number,Duan Ci,

Remaining time and number of cycles;

Accumulated operation time function;The temperature and humidity program settings are displayed as graphical curves,

Real time display of program curve execution function;Having a separate program editing screen,5Each page can be inputted

Temperature, humidity, and time of each segment;

Chinese and English can be switched freely;

Fault prompt screen display;

The screen can be adjusted for backlight;The screen display protection function can be timed,TIMER.

5Or manually disable the settings

Program capacity and control functionAvailable program groups:250*BigindividualPATTEN

.Available memory capacity:commontwelve thousand and five hundred

SEGMENTS.Repeatable command execution32000:

Each command can reachNext time.The production of the program adopts a conversational approach

,It has functions such as editing and insertion.0SEGMENTSTime setting~

540Hour59Min.Programmable timing control module device

x2Group.Having power-off program memory

,Automatically start and continue executing program functions after power restoration.utensil

RS-232Communication interface.utensil

USB

Interface functionality.

Real time display of graphical curves during program execution.

It has the function of automatically adjusting the freezing capacity.It has the function of scheduled startup and shutdown.Having a date

,Time adjustment function.Button and screen lock(

LOCK

1)Function.

**Protective device

2refrigeration system

Compressor overheat protection; Compressor overcurrent protection; Compressor overpressure protection; Overheating protection for condensing fan

3Chamber

Adjustable over temperature protection; Extreme over temperature protection for air conditioning channels; Fan motor overheat protection

other

Main power phase sequence and phase loss protection; Leakage protection; Heating, humidification, dry burning protection

1Overload and short circuit protection; power-off protection

Other configurationsSample power control terminalRelay contact control,AC220V、.

22A

Within (when operating normally, the contacts are closed; when the equipment stops or malfunctions, the contacts are disconnected)Main power supplyAC380V、

350Hz

、Power cableFive core (three-phase four wire)3+1Protective earth wire) cable

4rice

Article (to be provided upon customer request)Lead holeThe lead hole has a diameter of φ50mm、 φ80mm、 φ100mm、 φ

120mmThe specifications, location, and quantity can be customized according to user requirements, provided that the box structure allows and does not affect performance

5 USBinterfaceThe controller can download historical curves and data, and historical data can be downloadedExcelFormat saving,You can download the required data by date. Historical data

JPG

Save in image format, you can zoom in and out as needed, and then download.

1transport

2The test box is of integral type and can be transported as a whole

size

weight

Larger transportation dimensions (excluding packaging): refer to "external dimensions"

Larger transportation weight (excluding packaging): see "weight"

1Conditions of Use

The user guarantees the following conditions

Installation site

The ground is flat, well ventilated, and there is no strong vibration around the equipment

There is no strong electromagnetic field around the equipment

There are no flammable, explosive, corrosive substances or dust around the equipment

There is appropriate space for use and maintenance around the equipment, as shown in the following figure:The detailed spatial dimension requirements of this device can be found in the specific spatial dimension diagram.A : Not less than80cmB : Not less than80cmC

2: Not less than

120cm5environmental condition35Temperature:

℃~℃

Relative humidity: ≤85%Air pressure:86kPa

3~

106kPaPower supply conditions power supplyAC220 VThree-phase four-wire

+Protect the ground wire

Installed power:5.8 KWAllowable voltage fluctuation range:380AC(±10%

)50V0.5Allowable fluctuation range of frequency:(±

)4Hz

The grounding resistance of the protective grounding wire is less than

4Ω

Require users to configure corresponding capacity air or power switches for the equipment at the installation site, and this switch must be independently used for the equipment0Requirements for storage environment+45When the equipment is not working, the temperature of the environment should be maintained

℃~0Within ℃

When the ambient temperature is belowAt ℃, the water remaining in the equipment should be drained completely to prevent the water inside the pipeline from freezing and damaging the pipeline (only water-cooled type)Programmable temperature and humidity control experimental chamber
Maintenance
:

Regularly check whether the three-phase current of the power supply is balanced, whether the fuse of the power supply is intact, and strictly prevent phase loss operation and damage to the motor.

The normal operating temperature of this equipment should be below 43 ℃. When the equipment does not operate at the normal operating temperature, high voltage protection will occur and the unit operation will stop, which belongs to the system protection of the equipment, not a quality fault.

3. It is necessary to regularly check whether there is too much dust accumulation on the heat exchange fins of the outdoor unit. If so, timely dust accumulation should be carried out. Use a soft bristled brush to brush up and down along the direction of the fins. Be careful not to overwhelm the heat exchange fins and affect ventilation.

4. Clean the air filter every 15 days to avoid dust blocking and affecting ventilation.

Equipment refueling and refrigerant addition should be strictly controlled and operated by professional personnel to avoid damaging the machine.

8 When using, if the fan stops running, it must be stopped immediately for inspection and troubleshooting before continuing to use.