Sampling reference standard:
HJ2526-2012 "Technical Standard for Portable Dietary Oil Fume Detector"
GB18483-2001 "Emission Standards for Cooking Fumes in the Catering Industry"
DB11/1488 "Emission Standards for Air Pollutants from Catering Industry in Beijing"
SZDB/Z 254-2017 "Specification for Oil Fume Emission Control in the Catering Industry"
GB/T 16157-1996 "Determination of particulate matter and sampling methods for gaseous pollutants in exhaust from stationary sources"
1 Overview
Portable rapid oil fume detector(hereinafter referred to as the detector) is a high-precision portable integrated multifunctional oil fume detector, mainly used to collect the concentration of oil fume emissions in flues and chimneys. Stable performance, easy operation, compact and portable, stable flow rate, can directly test oil fume related data on site, greatly reducing labor intensity. Suitable for fixed flow sampling and concentration measurement of oil fume in fixed pollution sources, as well as temperature, humidity, dynamic pressure, static pressure and other working parameters and measurements of smoke, to evaluate the concentration of organized emissions of oil fume. It is widely used in restaurants, canteens, hotels, environmental monitoring stations, universities, research institutions, urban law enforcement teams and other places to detect the concentration and working parameters of oil fume emissions.
2 Main features
2.1adopt5-inch industrial grade high brightness color touch LCD display screen, graphic interface, easy and sensitive operation;
2.2The sampling point can be inputted using both Chinese and English input methods. At the end of the sampling process, the sampling location will be automatically saved to the data results. When querying the data, it is easy to know the sampling location of the current gas path sampling data;
2.3Real time monitoring of oil fume emission concentration in flue and chimney using laser scattering method, fully automatic measurement, high detection accuracy, good data repeatability, eliminating tedious manual sampling and analysis steps
2.4Real time measurement of smoke temperature, humidity, and other operating parameters
2.5The sampling time and save times can be customized to easily achieve monitoring tasks
2.6data storage300 sampled data can be viewed, printed, and exported from a USB drive;
2.7Built in large capacity lithium battery for continuous operation of the instrument, not less than5 hours;
3 Main Technologiesindicator
Main parameters |
parameter range |
resolution |
accuracy |
Oil fume concentration |
(0~50.0)mg/m3 |
0.01mg/m3 |
≤2mg/ m3Time is better than±0.2mg/m3, >2mg/ m3Time is better than±5% FS |
sampling flow |
4.0L/min |
------- |
---------- |
Smoke humidity |
(0~60.0)% |
0.01% |
superior to±2.0% |
flue gas temperature |
(-55.0~125.0)℃ |
0.1℃ |
superior to2℃ |
dynamic pressure |
(0~2000)Pa |
0.1Pa |
superior to±2.0%FS |
static pressure |
(-35~35)kPa |
0.01kPa |
superior to±4.0%FS |
atmospheric pressure |
(50 to 115 kPa |
0.01kPa |
not exceed±500Pa |
flue gas velocity |
(0~45)m/s |
0.01m/s |
superior to±5.0% |
Overall dimensions |
980mm×170mm×80mm |
||
Sampling tube length |
0.98 meters (customizable) |
||
Total power consumption |
6W |
||
Total weight |
<3.5kg |
||
working voltage |
Built in lithium battery orDC12.67V |
||
4.1 Working environment temperature: (-30~50)°C
4.2 Relative humidity of working environment: ≤ 85%
4.3 Atmospheric pressure: (80~105) kPa