The testing principle of the plastic particle moisture meter is the oven drying weight reduction method. The instrument automatically calculates the percentage of moisture content (moisture content) of the plastic particles based on the difference in mass of the plastic particle sample before and after heating and the percentage of the original mass before heating. The plastic particle moisture meter is several times or even dozens of times faster than the traditional oven drying method, and the measurement results of the plastic particle moisture meter are very close to the measurement data of the traditional oven drying method. The plastic particle moisture meter has a relatively simple measurement operation, does not require special training, and does not require manual calculation. It has gradually replaced the traditional oven method and become a testing material or product for many enterprises
Plastic particle moisture metertallPrecision imported weighing sensors and high-efficiency imported halogen lamp heating devices achieve fast and accurate moisture measurement. The halogen moisture analyzer is a reliable solution for measuring the moisture content of your product.
Applicable industries: widely used in grain, feed, tobacco, paper making, food (dehydrated vegetables, meat, noodles, flour, biscuits, pies, aquatic processing), tea, beverages, grains
In industries such as materials, chemical raw materials, pharmaceutical raw materials, textile raw materials, etc., free moisture content in samples is tested. advantage:
(1) High precision (imported sensors) combined with stability, easy to operate
(2) Powerful testing capabilities to obtain accurate data results
(3) LCD display screen, easy to read parameters
(4) Not affected by environment and humidity, no exposed auxiliary equipment
(5) Weight loss heating mode (standard heating, step heating, rapid heating analysis method) meets the needs of various solid, liquid, and powder moisture determination
(6 standard RS232 communication interface - convenient for connecting printers, PCs, and other peripheral devices. (7) The heating method of the circular quartz tungsten halogen infrared ring lamp can directly heat the material from the inside, greatly reducing the drying time. It also has uniform heating, cleanliness, high efficiency, and energy conservation. (The tungsten halogen ring lamp injects halogen gases such as iodine or bromine into the infrared lamp. At high temperatures, the sublimated tungsten wire reacts chemically with the halogen, and the sublimated tungsten will re solidify on the tungsten wire, forming a balanced cycle to avoid premature breakage of the tungsten wire. Therefore, tungsten halogen lamps are longer than ordinary infrared lamps.)
Plastic particle moisture meterComparison with oven method
ovenTest Report |
model |
oven1 |
XY-100MW |
Experiment Number |
1 |
2 |
3 |
4 |
1number |
test temperature |
105° |
105° |
Heating mode |
/ |
Standard heating |
Shutdown mode |
/ |
automatic shutdown |
operation steps |
weighing10gPlace the left and right samples in the oven105Baking degree2Weigh by hour and convert to moisture content. |
/ |
Weight before drying |
10.0794g |
10.0744g |
10.0812g |
10.0644g |
|
dryafterweight |
10.0172g |
9.9911g |
9.9707g |
9.998g |
9.9992g |
Test time |
2hour |
2hour |
2hour |
2hour |
4divide54second |
moisture content |
0.617%m |
0.826%m |
1.096%m |
0.659%m |
0.041%m |

ovenTest Report |
model |
oven2 |
XY-100MW |
Experiment Number |
1 |
2 |
3 |
4 |
2number |
test temperature |
105° |
105° |
Heating mode |
/ |
Standard heating |
Shutdown mode |
/ |
automatic shutdown |
operation steps |
weighing10gPlace the left and right samples in the oven105Baking degree2Weigh by hour and convert to moisture content. |
/ |
Weight before drying |
10.0587g |
10.0082g |
10.0508g |
10.0567g |
|
dryafterweight |
9.6648g |
9.6885g |
9.744g |
9.7151g |
9.9992g |
Test time |
2hour |
2hour |
2hour |
2hour |
4divide54second |
moisture content |
3.916%m |
3.194%m |
3.052%m |
3.396%m |
0.041%m |