Phone
(0)13809890260
-
Address
Room 2503, Block C, Huangdu Plaza, No. 3008 Yitian Road, Futian District, Shenzhen
Shenzhen York Technology Co., Ltd
(0)13809890260
Room 2503, Block C, Huangdu Plaza, No. 3008 Yitian Road, Futian District, Shenzhen
Performance indicators:
|
measurement range
|
Temperature -10-90 ℃ (or according to customer requirements)
|
|
Humidity 10-98% RH
|
|
|
measurement accuracy
|
temperature+0.5℃
|
|
humidity+3%RH
|
|
|
power supply
|
18-36VDC
|
|
output signal
|
Temperature 4-20mA (corresponding to -10-90 ℃)
|
|
Humidity 4-20mA (corresponding to 0-100% RH)
|
|
|
load resistor
|
R<500Ω
|
|
Equipped with display transmitter
|
Temperature range 0-80 ℃
|
|
display resolution
|
Temperature 0.1 ℃
|
|
Humidity 1% RH
|
|
|
character height
|
12.7mm
|
Dimensions (mm):
Maintenance: The probe should not be used in an environment with organic solvent vapors such as acetone. The dust cover can prevent airborne dust and fiber contamination of the humidity sensor in the gas. If blockage is found in the micropores of the dust cover, the cover can be removed, cleaned, and dried for use. Temperature fluctuations should be prevented during use to prevent probe condensation, which can affect measurement accuracy and corresponding time.
When installing the transmitter, it is not advisable to directly face the probe towards strong airflow or ventilation duct outlet. The probe should withstand airflow velocity of less than 15 meters per second. If there is mechanical vibration in the installation location, consideration should be given to adding shock absorbers.