The PH composite electrode CR 5503 is specifically designed for use in extremely harsh environments, and can be installed in pipelines or immersed installations. The reference electrode is made of a dual structure, which can extend the service life of the electrode in harsh applications. The electrode body material with threaded structure is corrosion-resistant Ryton (PPS), and the reference structure is porous Teflon or ceramic.

Product Details

PH composite electrode CR 5503 technical parameters
PH range: 0-14PH
ORP range: ± 1500mv
Temperature range: 0-105 ℃
*High pressure temperature: 150 PSI at 100 ℃
Reference electrode: Ag/AgC1
Liquid connection material: Ryton. Teflon, glass
Standard cable length: 10 feet
*High flow rate: 10 feet (3 meters) per second
Output impedance:<250 megaohms at 25 ℃
Corrosion resistance: These high-quality electrode materials are highly corrosion-resistant,
There is Ryton @, Teflon @ and Glass

American PH electrode, detailed introduction of Broadley James
1. American pH electrode, broadley james, Manufacturer BROADEEY JANEC, discharge outlet PH electrode, chrome tank PH electrode, cyanide tank PH electrode 1. Low internal resistance, therefore less interference

2. Using circular ceramic holes with a relatively large contact area, it is not easy to block the ceramic holes. The internal structure design adopts a stepped partition design to prevent external high concentration (Sample) infiltration, which can withstand high pollution

3. The anti UV HT-3 special blue glass head design detection system can effectively prevent Ag/AgCl polarization caused by UV irradiation in the glass detection system, increasing the service life of the electrode

Applicable occasions: Pure water/wastewater detection system

Measurement range: 0-14PH
Temperature range: 0-100 ℃

Voltage resistance: 0.6MP
Installation size: Upper and lower 3/4NPT pipe thread

texture of material: PPS
Connection method: Low noise cable directly led out