Welcome Customer !

Membership

Help

Jilin Puretech Technology Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Products

PCK208A Intelligent Open Flash Point and Ignition Point Tester

NegotiableUpdate on 05/02
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

The PCK208A intelligent open flash point and ignition point tester is designed based on the relevant provisions of the "National Standard of the People's Republic of China - Determination of Flash Point and Ignition Point of Petroleum Products - Cleveland Open Cup Method GB/T 3536-2008", "ASTM D92", and "ISO2592-2000", using microprocessor control technology combined with a true color touch LCD display system ..

Product Details

 
PCK208A Intelligent Open Flash Point and Ignition Point Tester
Overview
The PCK208A intelligent open flash point and ignition point tester is designed based on the relevant provisions of the "National Standard of the People's Republic of China - Determination of Flash Point and Ignition Point of Petroleum Products - Cleveland Open Cup Method GB/T 3536-2008", "ASTM D92", and "ISO2592-2000", using microprocessor control technology combined with a true color touch LCD display system.
feature
Chinese English bilingual menu, true color GUI touch screen interface design, intuitive display, easy operation (under certain conditions, it is completely possible to quickly familiarize oneself with the operation process without reading the manual), reasonable instrument structure, wide testing range, good repeatability and authenticity, etc. The testing process of this instrument is fully automated, and the temperature, time, and other parameters involved are displayed on the LCD screen. On this LCD screen, users will also see the working status of the instrument and related operation prompts at the same time.
You only need to preset the expected flash point of the substance to be tested, and the instrument can heat up, scan, detect the flash point, detect the flash point, extinguish the fire, lift the detection device, cool down, print and output the results at the specified rate, eliminating the complex operation process that you had to manually operate in the past; It has both automatic atmospheric pressure detection and manual input functions. When the automatic detection status is valid, manual input is invalid, and vice versa. The instrument will automatically adjust the test results based on changes in atmospheric pressure.
This instrument has powerful data processing capabilities: it can remember the results of 100 previous analyses, making it convenient for you to query and print historical records. Can remember the preset parameters before each shutdown. This instrument comes with a calendar that can display year, month, day, hour, minute, and second. Optional 232 computer management interface (one set of device computer software)
Instrument structure
The instrument is designed as a single unit, with beautiful appearance, compact structure, light weight, and small volume.
The control core is equipped with functional circuits such as power supply, interface, clock, driver, LCD display system, as well as executing components such as heating electric furnace, fire extinguishing mechanism, detection mechanism, gas source control valve, and lifting mechanism to complete the automatic control of the entire inspection process.
indicator parameters
Execute standard GB/T3536-2008 ASTM D92 ISO2592-2000
Temperature measuring element platinum resistance (Pt100)
Temperature range: Room temperature -400 ℃
Resolution 0.1 ℃
Measurement repeatability ± 4 ℃,
The heating rate is 14 ℃ -17 ℃/min before the pre flash point of 56 ℃,
Starting from 23 ℃ before the pre flash point, the temperature ranges from 5 ℃ to 6 ℃/min
Display method: True color touch LCD display
The usage environment has an ambient temperature of 15-35 ℃ and an ambient humidity of less than 85%
Power requirement 220V. ac ± 10% 50Hz ± 1Hz
Power loss ≤ 600W
Maximum size 480 * 290 * 220
Instrument weight 20Kg
PCK208A Intelligent Open Flash Point and Ignition Point Tester
Overview
The PCK208A intelligent open flash point and ignition point tester is based on the 'National Standards of the Peoples Republic of China-Determination of the Flash Point and Fire Point of Petroleum Products-Cleveland Open Cup Method-GB/T 3536-2008', 'ASTM D92', 'ISO2592- The relevant regulations of 2000 are designed using microprocessor control technology combined with a true color touch LCD display system.
Features
Chinese and English bilingual menu, true color GUI touch screen interface design, intuitive display, easy operation (under certain conditions, you can quickly familiarize yourself with the operation process without reading the manual), the instrument structure is reasonable, the test range is wide, the repeatability and authenticity are good and many more. The test process of the instrument is fully automatic, and the involved temperature, time and other parameters are all displayed on the LCD. On this LCD screen, the user will also see the working status of the instrument and related operation prompts at the same time.
You only need to preset the expected flash point of the substance to be tested, and the instrument can heat up, sweep, flash point detection, flash point detection, extinguishing (manual and automatic optional), detection device upgrade, air cooling, Printing and outputting results, etc., eliminates the complicated operation process that you had to do by yourself in the past; it has both automatic atmospheric pressure detection and manual input functions, and manual input is invalid when the automatic detection status is valid, and vice versa. The instrument will automatically correct the test results according to atmospheric pressure changes.
The instrument has powerful data processing functions: it can memorize the past 100 analysis results to facilitate your query and printing of historical records. The preset parameters before each shutdown can be memorized. The instrument comes with a hundred-year calendar, which can display year, month, day, hour, minute, and second. Optional 232 computer management interface (a set of equipment computer software)
Instrument structure
The instrument is a single-piece design with beautiful appearance, compact structure, light weight and small size.
The control core is equipped with functional circuits such as power supply, interface, clock, drive, liquid crystal display system, and execution components such as heating electric furnace, fire extinguishing mechanism, detection mechanism, air source control valve and lifting mechanism to complete the automatic control of the entire inspection process.
index parameter
Executive standard    GB/T3536-2008 ASTM D92 ISO2592-2000
Temperature measuring element      platinum resistance (Pt100)
Temperature range      room temperature---400℃
Resolution       0.1℃
Measurement repeatability     ±4℃,
Heating rate      before the pre-flash point 56℃ is 14℃-17℃/min,
Starting from 23℃ before the pre-flash point, it is 5℃-6℃/min.
Display mode      true color touch LCD display
Use environment      Ambient temperature 15--35℃ Ambient humidity is less than 85%
Power requirements      220V.ac±10% 50Hz±1Hz
Power loss    ≤600W
Maximum size       480*290*220
Instrument quality         20Kg