PAMAS SBSS desktop particle counter

NegotiableUpdate on 06/07
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

PAMAS SBSS desktop particle counter

Product Details

Detailed introduction:

principle
SBSS
The particle counting system used in the laboratory adopts injection bottle sampling, which is mainly used for measuring the pollution level of solid particles in oil, polymer solutions, and gel solutions. Based on the principle of light blocking, a single particle counting system is designed and manufactured. Due to the use of specially manufactured sensing cells and the selection of precision optical components, high resolution and accuracy can be ensured. Each particle detected by the sensor will be counted according to its size. Due to the setting of up to32A particle size channel that can be adjusted freely, thus displaying a complete particle number and size distribution map. Even for demanding application scenarios, it can still provide a barrier between0.5-800Sensors used in different size ranges between micrometers can automatically sample and report data from multiple samples using an analysis program controlled by a programmable processor.


Applicable samples
--
Standard particle pollution control (pressure not exceeding)10Ba, viscosity not greater than400Li Si)
Hydraulic oil, insulation oil and turbine oil, gear oil and engine oil, water-based hydraulic oil, infusion solution, organic liquid, sensitivity1Microns(ISO 4402)Sensitivity4Microns(ISO 11171
--
Special purpose particle pollution control (pressure not exceeding)30Ba, viscosity not greater than20,000Li Si)
Oil, polymer solution, gel like system

Particle counter calibration
SBSS
The Type Particle Counter is a single type particle counting system with dual calibration function, which can be used according toISO-11171as well asISO-4402After one calibration, it can be used,ISO-4406The third level code can be printed in the following format:
according toISO-11171The calibration method4Micrometer,6Microns and14Micron code system.
according toISO-4402The calibration method2Micrometer,5Microns and15Micron code system.

Classification function of any standard
Within the particle size range of the sensor, any national or international standard (e.g,ISO-4406NAS1638SAE749dASTMClassify the number and size of particles.

Sensor calibrate
Sensor calibration based onISO-11171as well asISO-4402conduct. Calibrate the pulse height using microemulsion standards according toANSI/NFPA T2.9.6R1-1900ASTM F658-80Calibration can be carried out according to personal standards, or based onUSP XXIVPerform calibration.
PAMAS SBSS
The pre adjustment automatic sample vacuum system
Up to32A particle size channel that can be adjusted freely.
Threshold adjustment passedD/AThe converter is being used.
Vacuum sampling (pressure up to)10Ba).
High pressure sampling (pressure up to30Ba).
Sample volume:50-1000milliliter.
LV
Container(2000Milliliters).
Analysis volume:1-1000milliliter.
The constant current and volume control of the sample are achieved through the adjustable movement of the piston under pressure.

Technical data of the counter
The point resolution is320x240Background illuminated LCD graphic display
Dot printing:32Built in thermal printer
Data transmission:8byteASCIICode andRS-232Cinterface
Power Supply:110V115V220V230V50-60Hz

Technical data of sensors
HCB-LD-50/50
type
Particle size range:1.0-450micron
Sensitivity:1Microns(ISO-4402
Sensitivity:4Microns(ISO-11171
Large particle concentration: in25ml/minute*When is24000particle/ml**
(Note:* You can choose different flow rates;** Accidental error less than7.8%

HCB-LD-25/25TypeHCB-LD-15/25TypeHCB-LD-100Type (omitted)