Welcome Customer !

Membership

Help

Shenzhen Huachang Haitai Technology Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Products

Shenzhen Huachang Haitai Technology Co., Ltd

  • E-mail

    szhcht@163.com

  • Phone

    18818683927

  • Address

    Room 556, Jinda Technology Center, No. 8 Kefeng Road, Nanshan District, Shenzhen

Contact Now

PAMAS S40 portable particle counter

NegotiableUpdate on 06/07
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

PAMAS S40 portable particle counter

Product Details

Detailed introduction:

principle
S40
The portable particle counting system for on-site use adopts a built-in piston pump, which can handle non pressure samples and pressures up to420The pressure system of Ba delivers a constant flow rate sample to the sensor, mainly used for measuring the degree of solid particle contamination in oil and other liquids. A single particle counting system designed and manufactured based on the principle of light blocking, using specially manufactured sensing cells and selecting precision optical components, can ensure high resolution and accuracy. Each particle detected by the sensor will be counted according to its size. set up8A particle size channel that can be adjusted freely and displays the complete number and size of particles.

characteristic
It has both portability and accuracy
Using touch screen operation, it can display the number of particles and graphics
Convenient software downloads
Compact standardized interface(DIN 2353
Built in battery power for data backup
Can be stored500Secondary measurement results
No pressure sampling or under pressure up to420Bashi sampling
When the viscosity does not exceed160The measurement result of Li Si Shi is independent of viscosity (the upper limit of viscosity during pressurized sampling is300Li Si)
Set password protection based on user level

Widely applicable scenarios
*
High pressure sampling (pressure up to420Ba)
*
Sensitivity:1Microns(ISO-4402
Sensitivity:4Microns(ISO-11171
* 8
Channel particle counter
*
Display Level 3 Code:2Micrometer,5Microns and15Microns(ISO-4402
4
Micrometer,6Microns and14Microns(ISO-11171andNASCleanliness Classification
*
Built in battery
*
Highly repetitive and accurate
*
Threshold adjustment passedD/AConverter (front panel or software)

Classification function of any standard
Within the particle size range of the sensor, the number and size of particles can be classified according to any national or international standard and the results can be printed out. For example, according toISO-4406NAS1638SAE749dASTMMIL-1246CPrinting in waiting format. You can also use convenient download software to transfer data to different computing programs.

Sensor calibrate
according toISO-4402useACFTDperform
according toISO-11171useISO MTDNIST)Conducted in oil. Use micro lotion standard to calibrate pulse height according toANSI/NFPA T2.9.6R1-1900andASTM F658-8conduct.

Technical data of the counter
* 8
Channel particle counter, particle size is defined in any of the following ways.
2
Micrometer,5Micrometer,10Micrometer,15Micrometer,20Micrometer,25Micrometer,50Micrometer,100Microns(ISO-4402);
4
Micrometer,5Micrometer,6Micrometer,10Micrometer,14Micrometer,20Micrometer,25Micrometer,30Microns(ISO-11171).
*
Data printing format:32Built in thermal printer
*
Data transmission:8byteASCIICode andRS-232CInterface
*
Power Supply:110V115V220V230V50-60Hzperhaps14.4VBattery(12-30VCommunication input)

PAMASTechnical data of capacitive sensors
HCB-LD-50/50
type
Particle size range:1-400micron
1-100
Microns(ISO-4402:1991
4-30
Microns(ISO-11171:1999
4-70
Microns(ISO-11171:1999
Large particle concentration: in25milliliter/In minutes, it is20000particle/milliliter
sampling system Constant flow piston pump, with viscosity not exceeding160The measurement result of Li Si Shi is independent of viscosity (the upper limit of viscosity during pressurized sampling is300Li Si).
Pressure Range0-7Ba,3-210Ba (maximum value)420Ba