Ø Key Features:
l Large screen dot matrix LCD 192 × 64 (or 320 × 240), multi parameter display, rich content.
l The adoption of advanced embedded system design and surface mount technology has improved
Product performance and reliability meet EMC design requirements.
l Chinese and English bilingual programmable switching to meet different user needs.
l Fully guided operation mode in both Chinese and English, easy to use, and easy to understand.
l Programmable automatic or manual temperature compensation method, flexible and convenient to use.
l Using high-performance composite electrodes for accurate measurement and rapid response.
l Two completely isolated current signals can be output, and the output current range can be set separately.
l Equipped with upper and lower limit alarm functions, the alarm values can be set separately.
l Equipped with a standard 485 digital communication interface, it can achieve long-distance communication.
l It has the functions of storing and querying historical data, operation and calibration records.
l Can query 100000 historical data, 1000 operational records, and 100 calibration records.
l High protection level, reaching IP66, can meet various complex environmental application requirements.
l The brand new circulation pool design has a simple structure and strong anti-interference ability.
Ø Main Technical Performance:
Display: Chinese and English display, 192 × 64 (or 320 × 240) dot matrix LCD
Measurement range: (0.00~14.00) pH
Resolution: 0.01pH
Instrument reading error: ± 0.05pH
Input impedance: not less than 1 × 1012 Ω
Response time: (T90) 90s (25 ℃)
Temperature sensor: Pt1000
Temperature range: (0.0~99.9) ℃
Temperature accuracy: ± 0.5 ℃
Temperature resolution: 0.1 ℃
Temperature compensation: (0.1-60.0) ℃ (manual or automatic)
Sample conditions: Temperature range: (5-50) ℃
Flow range: no more than 6 liters per hour
Environmental temperature: (5-45) ℃
Environmental humidity: not exceeding 90% RH (without condensation)
Current output: (4-20) mA (two-way isolated output)
Current accuracy: ± 1% FS
Current load:<800 Ω
Alarm output: two-way alarm output, DC 5A/30V or AC 5A/250V
Storage and transportation temperature: (-20~55) ℃ (excluding electrodes, electrodes should be above 0 ℃)
Dimensions: 144mm x 144mm x 210mm (width x height x depth)
Hole size: 139mm × 139mm
Power supply: AC (85-265) V, frequency (45-65) Hz
Power: ≤ 10W
Weight: Approximately 1.5 kg