The oil particle counter is based on standards such as GB/T 18854-2002 (ISO11171-1999). Developed using the principle of light blocking counting, in compliance with relevant national and international standards. It can provide fast, accurate, reliable, and reproducible detection results and complete pollution monitoring and analysis reports, suitable for detecting particle pollution levels in hydraulic oil, lubricating oil, fire-resistant oil, insulation oil, and turbine oil. It can be widely used in aviation, aerospace, power, petroleum, chemical, transportation, ports, metallurgy, machinery, automobile manufacturing and other fields.

Product Details

Oil particle counterMain Features


★ Adopting the counting principle of photoresist (shading) method developed by the International Hydraulic Standards Committee

★ High precision laser sensor with a wide testing range, stable performance, low noise, and high resolution

★ Adopting high-pressure injection pump sampling method, the sampling volume can be set by oneself, the injection speed is stable, and the sampling accuracy is high

★ The injection system combining positive and negative pressure can achieve sample degassing and is suitable for testing samples with different viscosities

★ Built in pressure sensor, capable of setting pressure values and automatically determining cabin pressure to ensure safety

★ Built in air purification system ensures that testing is not contaminated

★ Built in multiple calibration curves, compatible with all commonly used domestic and international standards for calibration

★ built-inGJB-420A、GJB-420B、NAS1638、GB/T14039、ISO4406、SAE4059cpc、SAE4059F、SAE749D、ГOCT17216、QC/T29104、JB/T9737、DLT432、HH005-2018 Waiting for more than ten commonly used standards, one test can provide data results for all built-in standards

★ Support custom standard testing and level determination

★ According to customer needs, up to simultaneous settings can be made64 detection channels

★ Built in data analysis system, one test can provide all standard test data and pollution levels

★ Built in viscosity, moisture, and temperature sensor modules provide accurate testing of particle distribution while providing viscosity, moisture saturation, andPpm value and temperature reference value (optional)

★ The particle size can be set arbitrarily,built-in9999Particle size, facilitating particle size analysis

★ Multiple sampling containers such as standard sampling bottles or cups can be used to meet the testing requirements of different industries

★ Fully functional automatic color touch screen operation, supporting both Chinese and English interfaces, easy and convenient to operate

★ Can perform single channel and multi-channel calibration to achieve automatic calibration function

★ haveRS232 interface, can be connected to a computer or laboratory platform for data processing

★ Equipped with massive data storage and printing functions, capable of storing1000 sets of data and support USB storage of data

Oil particle countertechnical indicators

  • Light source: semiconductor laser

  • Particle size range:0.8 μ m~600 μ m (calibrated according to 1-100um or 4-70um (c))

  • Detection channel:8-64 channels are optional, with any particle size set

  • Sensitivity:0.8μm(ISO4402) 或3 μm(c)(GB/T18854,ISO11171)

  • Resolution:<10%(GB/T18854,ISO11171)

  • Repeatability:RSD<2%

  • Sample viscosity testing:≤ 650cSt (excessive viscosity can be detected by heating or dilution methods)

  • Sampling volume:0.2~1000mL, Interval of 0.1mL

  • Sampling accuracy: better than±0.5%

  • Sampling speed:5 ~80mL/min

  • Pressure chamber: positive and negative pressure chamber device, realizing sample degassing and high viscosity sample detection

  • Maximum vacuum in the pressure chamber:-0.08Mpa

  • Maximum positive pressure in the pressure chamber:0.8 Mpa

  • Maximum particle concentration:12000~40000 particles/mL

  • power supply110~245V AC、50Hz,70W

  • Overall dimensions:340mm╳410mm╳650mm