- Phone
-
Address
Room 325, Building D-1, Zone D, Service Outsourcing Industrial Park, No. 8 Xinfeng Road, Daqing High tech Zone, Heilongjiang Province
Daqing Dongding Instrument Manufacturing Co., Ltd
Room 325, Building D-1, Zone D, Service Outsourcing Industrial Park, No. 8 Xinfeng Road, Daqing High tech Zone, Heilongjiang Province
Oil particle counterMain Features
★ Adopting the counting principle of photoresist (shading) method developed by the International Hydraulic Standards Committee
★ High precision laser sensor with a wide testing range, stable performance, low noise, and high resolution
★ Adopting high-pressure injection pump sampling method, the sampling volume can be set by oneself, the injection speed is stable, and the sampling accuracy is high
★ The injection system combining positive and negative pressure can achieve sample degassing and is suitable for testing samples with different viscosities
★ Built in pressure sensor, capable of setting pressure values and automatically determining cabin pressure to ensure safety
★ Built in air purification system ensures that testing is not contaminated
★ Built in multiple calibration curves, compatible with all commonly used domestic and international standards for calibration
★ built-inGJB-420A、GJB-420B、NAS1638、GB/T14039、ISO4406、SAE4059cpc、SAE4059F、SAE749D、ГOCT17216、QC/T29104、JB/T9737、DLT432、HH005-2018 Waiting for more than ten commonly used standards, one test can provide data results for all built-in standards
★ Support custom standard testing and level determination
★ According to customer needs, up to simultaneous settings can be made64 detection channels
★ Built in data analysis system, one test can provide all standard test data and pollution levels
★ Built in viscosity, moisture, and temperature sensor modules provide accurate testing of particle distribution while providing viscosity, moisture saturation, andPpm value and temperature reference value (optional)
★ The particle size can be set arbitrarily,built-in9999Particle size, facilitating particle size analysis
★ Multiple sampling containers such as standard sampling bottles or cups can be used to meet the testing requirements of different industries
★ Fully functional automatic color touch screen operation, supporting both Chinese and English interfaces, easy and convenient to operate
★ Can perform single channel and multi-channel calibration to achieve automatic calibration function
★ haveRS232 interface, can be connected to a computer or laboratory platform for data processing
★ Equipped with massive data storage and printing functions, capable of storing1000 sets of data and support USB storage of data
Oil particle countertechnical indicators
Light source: semiconductor laser
Particle size range:0.8 μ m~600 μ m (calibrated according to 1-100um or 4-70um (c))
Detection channel:8-64 channels are optional, with any particle size set
Sensitivity:0.8μm(ISO4402) 或3 μm(c)(GB/T18854,ISO11171)
Resolution:<10%(GB/T18854,ISO11171)
Repeatability:RSD<2%
Sample viscosity testing:≤ 650cSt (excessive viscosity can be detected by heating or dilution methods)
Sampling volume:0.2~1000mL, Interval of 0.1mL
Sampling accuracy: better than±0.5%
Sampling speed:5 ~80mL/min
Pressure chamber: positive and negative pressure chamber device, realizing sample degassing and high viscosity sample detection
Maximum vacuum in the pressure chamber:-0.08Mpa
Maximum positive pressure in the pressure chamber:0.8 Mpa
Maximum particle concentration:12000~40000 particles/mL
power supply110~245V AC、50Hz,70W
Overall dimensions:340mm╳410mm╳650mm
Net weight of the instrument:25kg