- Phone
-
Address
No. 193 Huixin Road, Nanhu District, Jiaxing City, Zhejiang Province
Beijing Huaxia Sci Tech Innovation Instrument Co., Ltd
No. 193 Huixin Road, Nanhu District, Jiaxing City, Zhejiang Province
1. Compliant with national standards
National Environmental Protection Standards of the People's Republic of China“HJ970-2018”Water quality Determination of petroleum products UV spectrophotometry (trial)
2. Using reagents:
hexane
3. Technical principles
existpH≤2Under certain conditions, the oil substances in the sample were extracted with n-hexane. The extract was dehydrated with anhydrous sodium sulfate and then adsorbed with magnesium silicate to remove polar substances such as animal and plant oils225nmMeasure absorbance at wavelength, petroleum content and absorbanceThe value conforms to Lambert Beer's law.
4. application area
Used for the determination of petroleum in surface water, groundwater, and seawater
5. Technical indicators and related parameters
●Test wavelength:225nm
●Wavelength accuracy: ±0.5nm
●resolving power:0.001 mg/L
●Measurement range:0-50mg/L
●Method detection limit: When the sample volume is500mL,The volume of the extraction solution is25mL, using2cmWhen using a quartz colorimetric dish, the detection limit is0.01mg/L.
●Instrument detection limit:DL<0.04mg/L(Determination of n-hexane blank solution)11Secondary3timesSD)
●Minimum detectable concentration:0.002 mg/L(Concentration in water sample)
●Accuracy error: ≤±2%
●Repeatability:RSD≤2%
●Linear correlation coefficient:r> 0.999
● Baseline stability:Real time automatic adjustment of zero point (eliminating the impact of baseline drift);
● Light source system:The light source uses deuterium lamps with high wavelength accuracy, and the lifespan can reach5000More than hours
Calibration method: Standard curve calibration
● Temperature Degree:-5-45℃
● Relative humidity:5%-95%
● Host power supply power:(220±22)V、(50±1)Hz、50VA
● Shell material: Overall engineering plastic structure(PC+ABS)Anti electromagnetic radiation treatment, impact resistance, anti-corrosion
● Weight:3kg
● Dimensions:320mm(long)×300mm(wide)×170mm(tall)
Concentration limits for petroleum in some environmental quality standards and emission standards in China
|
Standard Name |
Standard number |
Concentration limit (unit:mg/L) |
|||||||||
|
environmental quality standards for surface water |
GB3838-2002 |
class i |
class ii |
class iii |
Class IV |
Class V |
|||||
|
≤0.05 |
≤0.5 |
≤1.0 |
|||||||||
|
Sea water quality standard |
GB3097-1997 |
class i |
class ii |
class iii |
Class IV |
||||||
|
≤0.05 |
≤0.30 |
≤0.50 |
|||||||||
|
Standards for drinking water quality |
GB5749-2006 |
0.3 |
|||||||||
|
Water quality standard for fisheries |
GB11607-89 |
≤0.05 |
|||||||||
|
Standards for irrigation water quality |
GB5084-52 |
paddy |
dry farming |
vegetable |
|||||||
|
≤0.50 |
≤10.0 |
≤1.0 |
|||||||||
|
Environmental Quality Assessment Standards for Edible Agricultural Products Production Areas |
HJ332-2006 |
paddy |
dry farming |
vegetable |
|||||||
|
≤0.50 |
≤10.0 |
≤1.0 |
|||||||||
|
Environmental Quality Evaluation Standards for Greenhouse Vegetable Production Areas |
HJ333-2006 |
≤1.0 |
|||||||||