Welcome Customer !

Membership

Help

Shanghai Shoufang Electronic Technology Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Products

Shanghai Shoufang Electronic Technology Co., Ltd

  • E-mail

    8200@shsoly.com

  • Phone

    17317182621

  • Address

    Room 307, Huacheng Plaza, 1471 Gumei Road, Minhang District, Shanghai

Contact Now

Murakami Color MCRL Angle Spectrophotometer from Japan

NegotiableUpdate on 01/20
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
Murakami Color MCRL Angular Spectrophotometer $r $n Model GCMS-11 $r $n Ternary Variable Angle Spectrophotometer $r $n Dual Beam Method Three Dimensional Variable Angle Spectrophotometer $r $n
Product Details

Murakami Color MCRL Angle Spectrophotometer from Japan

Model GCMS-11

Tri variable angle spectrophotometric system

Dual beam method three-dimensional variable angle spectrophotometer system


In terms of measuring the trigger effect of pearl mica coating, grating tiles, colored stainless steel, etc. based on color changes when the observation direction and irradiation direction change. Using a direct drive motor4The automatic angle changing mechanism can quickly process a large number of three-dimensional angle changing spectral color measurements when the axis works simultaneously (except for transmission color settings). Transmission color can also be measured.

Variable angle axis4Axis automatic control corresponding

The use of variable angle parametersLCDThe conventional coordinate system (incident angle θ)i, incident azimuth angle φi, light receiving angle θrThe azimuth angle of the received light is φr). Because the device directly interacts with θi,φi,θr,φrThe mechanism for variable angle can be directly specified using θi,φi,θr,φrconduct.

JIS Z 8722Color measurement method-Reflection and Transmission of Object Colors4.33The measurement method of spectral solid angle reflectance conforms to the methoda(When using a dual path spectrometer for substitution). Its characteristic is that the incident beam and the received beam overlap, and the angle range that cannot be measured is as small as10.

There are also corresponding backlight sources availableLCDAnd organicELWaiting for the color of the self luminous light sourceGCMS-10Type.

Murakami Color MCRL Angle Spectrophotometer from Japan

Technical Specifications

Spectrophotometer Department

Color measurement method

Dual beam method

Spectrophotometer

Concave analysis grid

Photo detector

Silicon photodiode array 256Motoko

Measure wavelength range

380nm~780nm

Number of output wavelength intervals

10nmEvery year41wavelength

Band pass

about10nm

Wavelength accuracy

0.2nm

A/DConverter resolution

16Bit

Lighting Department Official

Composition of light source room

Lamp, elliptical mirror, cold filter, shutter, starter, stabilized power supply, fan for air cooling, etc

light source

High speed xenon lamp 150W

light guide

Illumination light/reference light2Branched optical fiber

Illumination optical system

Within the incident plane±2.2°(Japanese shape)

Variable Angle Mechanism Department

Number of angle axes

4(3Dimensional Corner Match)

Variable angle mechanism

Return inside the sample table(入射方位角φiJiao Jiao)Incident angleθiAngle of rotation, azimuth of light receptionφrCorner angle, light receiving angleθrJiao Jiao

Automating

Above4Automatic control

Range of angle variation of incident angle and light receiving angle

~77°(Except for the following)

Range of azimuth angle variation

~360°(The incident and light reception are also excluded from the following invariant range)

Corner cannot be domain

The angle of incidence of incident light is not limited to the angle of incidence which does not obstruct both luminous flux of incident light and light receiving optical axis10°Azimuth angle17°

Angle accuracy

±0.5°Internal position detector decomposition energy
614,400Pulse / rotation

*Pair positioning accuracy

150second

Iterative positioning accuracy

±2.1second

Measure area

Light receiving angleAboutφ377°About3×13.3(楕 yen)

Light receiving optical system

±2.2°(Japanese shape)

Sample table

samplesize

165×130within(Radius from center of rotation axis195In the circle)

Sample thickness

50within

ZAxis correction(Thickness)

Micro adjustment range10

X,YAxis fixation

Fixing with fixed fixture

X,Yaxis movement

Micro adjustment range±12.5

Correction of inclined measurement surface

alphaaxis±2Baxis±2

Xu Rong's Load

about4kg

Polarization degree, measurement efficiency, and accuracy

Polarization degree(P)

1.52below

Data reading time

1Per set3Within seconds

Repeatability

0.05(Standard white plate45゜ Incident-0When measured by light receiving560nmStandard deviation)