| National standards implemented | GB/T 3785-2010 (IEC 61672-1:2013) Level 1 GB/T 3241-2010 Level 1/IEC 61260-1:2014 Class 1 |
| Execution of metrological verification regulations | JJG 188-2017 Sound Level Meter, JJG 778-2005 Noise Statistical Analyzer, JJG 449-2014 Octave and Fractional Octave Filters |
| Other compliant standards | IEC60651 Class 1,IEC60804 Type 1,IEC61252 ,GB/T 15952 ,ANSI S1.4A-195 Type1,ANSI S1.43 Type1 , ANSI S1.11-2004 Class 1,ANSI S1.25-1991,DIN 45675(1997-07) JJF 1555-2015 |
| frequency range | 10 Hz~20 kHz |
| Measurement upper limit | Low range: 132 dB (A), high range: 142 dB (A), customizable 162 dB (A) |
| Self generated noise | Low range: less than 12 dB (A), 17 dB (C), 22dB (Z) High range: less than 23 dB (A), 27 dB (C), 32 dB (Z) |
| Linear range | Greater than 112 dB (A) |
| Frequency weighting | Parallel (simultaneous) A, C, Z 1/3 OCT with B, D, T (custom), U (custom) weighting |
| Time weighting | Parallel (simultaneously) F, S, I, and Peak |
| indicator | 2.6-inch color screen display, resolution 240 × 320, rich display content. When the backlight delay is normally on, the brightness is automatically adjusted, and others can be manually adjusted |
| Main measurement indicators | Lxyi、Lxyp、Lxeq、Lxmax、Lxmin、LxN、SD、SEL、LCpeak 等 Note: x is A, C, Z, y is F, S, I, N are 1-99 integers that users can choose from |
| Peak C-level measurement range | Low range: 50 dB~132 dB High range: 60 dB~142 dB |
| Integral measurement time | Manual, set freely or in stages from 1 second to 99 hours |
| A/D sampling frequency | 48k times/second |
| 24-hour automatic monitoring | Measure once per hour, with each measurement time ranging from 1 minute to 1 hour, and can continuously measure multiple groups for 24 hours |
| Real time parallel analysis of filters | 1/1倍频程: 16 Hz、31.5Hz、63Hz、125Hz、250Hz、500Hz、1kHz、2kHz、4kHz、8kHz、16 kHz |
| 1/3倍频程: 12.5 Hz、16Hz、20Hz、25Hz、31.5Hz、40Hz、50Hz、63Hz、80Hz、100Hz、125Hz、160Hz、200Hz、250Hz、315Hz、400Hz、500Hz、630Hz、800Hz、1000Hz、1250Hz、1600Hz、2000Hz、2500Hz、3150Hz、4000Hz、5000Hz、6300Hz、8000Hz、10000Hz、12500Hz、16kHz、20kHz | |
| FFT analysis function | 2048 line, sampling frequency: 48kHz, 24kHz, 12kHz, 6kHz, 3kHz optional |
| FFT analysis window function | Rectangular window, Brinell window, Hanning window, flat top window |
| store | Equipped with 3 MB as standard, it can store up to 2663 sets of statistics, 3328 sets of integrals, 3328 sets of OCT, 1664 sets of 1/3 OCT or 121 sets of FFT analysis results, or can record up to 34 hours of personal sound exposure data in a single session, for a total of up to 320 hours |
| output interface | 1) Communication output, output power: 150 mW, can be connected to 8 Ω monitoring headphones 2) DC output: Output a DC signal proportional to the current displayed sound pressure level, 15 mV/dB 3) RS232 interface: communicates with a computer and can print measurement results and related charts through a related micro printer 4) USB interface: communicate with computer, firmware upgrade, compliant with USB1.1 standard, compatible with USB2.0 standard |
| Internal calendar clock | Error less than 1 minute/month, GPS timing and calibration available |
| power supply | 4 AA alkaline batteries, capable of continuous use for over 30 hours, meeting the requirement for 24-hour continuous monitoring in environmental monitoring |
| Dimensions (mm) | 260×80×30 |
| Mass (kg) | 0.35 |
| working environment | Working temperature range: -15 ℃ to 55 ℃; Relative humidity: 20% to 90%; Atmospheric pressure: 65 kPa to 108 kPa |
| SD card storage module (optional) | Maximum support for 32 GB |
| Recording function (optional) | Recording accuracy: optional 48kHz/32-bit and 8kHz/8-bit Trigger method: Synchronize with integral measurement or record the specified time after exceeding the limit. The longest continuous recording time is 1 hour at 48kHz/32-bit and 16 hours at 8kHz/8-bit. |
| calibration | Supporting the use of AWA6223F multi frequency point calibrator, meeting the requirements for sound level calibrators in HJ 707-2014 Environmental Sound Monitoring Technical Specification |
| Print function | AH40 dot matrix printer, printing data can be saved for a long time |
| GPS positioning (optional) | Internally embedded GPS positioning system, measuring noise while obtaining measurement point position and motion speed information |
| Bluetooth wireless transmission | Internally embedded Bluetooth module, sending measurement results to designated mobile phones or computers |
| place | Configuration Name | Remarks |
| host | AWA6228+host (including test microphone, preamplifier, etc.) | (Measurement software must be optionally selected) |
| Software selection | S6228+-00000 Basic measurement function | |
| S6228+-00101 Noise Statistical Analysis | ||
| S6228+-00202 1/1 OCT analysis software | ||
| S6228+-00303 1/3 OCT analysis software | ||
| S6228+-00404 FFT analysis software | ||
| S6228+-00013 DOSIMeter software | ||
| S6228+-00509 digital recorder (data acquisition and recording) | ||
| S6228+-00508 Reverberation Time Measurement Software | ||
| Hardware selection | GPS positioning module (built-in, including GPS antenna) | |
| SD storage module (built-in, including 32GB capacity SD card) | ||
| AWA8570A power module | ||
| AWA8570B SMS module | ||
| AWA8570AB GSM/solar cell module | ||
| AWA8732 type extension cable (5m, 10m, 20m) | ||
| Small tripod (including threaded fixture) | ||
| AWA2581 Outdoor Monitoring Box | ||
| AH40 printer (including 5 rolls of printing paper and instruction manual) |