Micellar EPR-IS Hydrogen Explosion proof Dew Point Sensor

NegotiableUpdate on 05/12
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

Micellar EPR-IS hydrogen explosion-proof dew point sensor Dew point sensors are designed to reliably and accurately measure dew point or moisture content in various gas or liquid analyzer applications. The compact mechanical design reduces installation time and provides a robust and reliable transmitter for all intrinsically safe hygrometer applications. The Easidew sensor can be used in conjunction with several moisture meters or moisture analyzers from Michell.

Product Details

Micellar EPR-IS Hydrogen Explosion proof Dew Point SensorIntended to reliably and accurately measure dew point or moisture content in various gas or liquid analyzer applications. The compact mechanical design reduces installation time and provides a robust and reliable transmitter for all intrinsically safe hygrometer applications. The Easidew sensor can be used in conjunction with several moisture meters or moisture analyzers from Michell.


Micellar EPR-IS Hydrogen Explosion proof Dew Point Sensor

.application field

  • Moisture in natural gas production and processing

  • Pipeline drying

  • Online financial measurement of natural gas

  • Dew point measurement in the production process of liquefied natural gas

  • Petrochemical process liquids

  • Measurement of Trace Humidity in Polymer Production

  • Moisture Analysis in CNG Drying

  • Dew point in the IS area

IECEx, cCSAus, FM, GOST, and ATEX dew point transmitters

Micellar EPR-IS Hydrogen Explosion proof Dew Point Sensor

technical indicators

product Easew PRO IS for gases Easew PRO LQ IS for liquids
performance metrics
measurement range -Dew point from 110 to+20 ° C; Dew point -100 to+20 ° C 0 to 1000 ppmW
accuracy ± 1 ° C dew point (+20 to -60 ° C)
± 2 ° C dew point (-60 to -110 ° C)
response time 5 minutes to T95 (dry to wet)
repetitiveness Dew point at 0.5 ° C
calibration Traceable 13 point calibration and certification
Electrical specifications
output signal 4-20 mA (2-wire connection, current source); User configurable range
output Dew point or moisture moisture content
Analog output scale range the dew point:-110 to+20 ° C or -100 to+20 ° CMoisture content in gas:0-3000 ppmV
non-standard: mg/m3, lbs/MMSCF natural gas
Moisture content in liquid:0 to 1000 ppmW
Non standard can be provided upon request
power supply voltage 14 to 28 V DC
load resistor 250 Ω @ 12 V (500 Ω @ 24 V)
Current consumption Up to 23 mA, depending on the output signal
Saturation constant
(Applicable only for measuring moisture in liquids)
6-point lookup table for achieving a saturation constant of 1000 ppmW within the temperature range of 0 to+50 ° C; The saturation constants of 10 common liquids can be programmed into Easew PRO XP IS through application software.
In addition, users can manually set the saturation constant

CE conformity Approved
Operating specifications
Operating Temperature -40 to+60 ° C
Compensation temperature range -20 to+50 ° C
Note: The transmitter accuracy statement is only valid within the following temperature range: -20/+50 ° C
Storage temperature -40 to+60 ° C
Operational pressure 高45 MPa(450 barg)
traffic 1 to 5 Nl/min, installed in a standard sampling block,
0 to 10 m/sec direct insertion
Using Easidew sample blocks
Directly insert 0.1 to 0.3L/min at a speed of 0.1 to 1m/s
Mechanical Specifications
Protection level IP66 complies with BS EN 60529:1992 standard;
NEMA 4 protection in accordance with standard NEMA 250-2003
Intrinsic Safety Zone Certificate ATEX:II 1 G Ex ia IIC T4 Ga(–20°C 至+ 70 °C)
IECEx:Exd ia IIC T4 Ga(–20°C 至+ 70 °C)
TC TR0Ex ia IIC T4 Ga(–20° C 到+ 70 ℃)
çCSAUS:Level I, my group ABCD, T4
FM:Level I, Zone 1, ABCD,T4
Russian pattern approval Russia (GOST-R), Kazakhstan (GOST-K)
Canadian Pressure Vessel Certification CRN-11 Canadian Provinces
Oxygen service Optional:Clean and rich in oxygen
Building materials 316 stainless steel
filter
(Sensor protection)
Standard:Stainless steel sintering protective cover (used to prevent fine particles>80 μ m)
Optional:HDPE protective cover (used to prevent fine particles>10 μ m)
Process connections and materials 316 stainless steel 5/8 inch -18 UNF, with adhesive seal, G1/2 inch and 3/4 inch UNF adapters
(Material certificate for BS EN 10204 3.1- please refer to "Accessories and Spare Parts" on the product ordering code)
weight 1.27 kilograms
electrical connection Screw terminal with internal thread M20 x 1.5mm cover
Diagnostic conditions (factory programming) stateoutput
Sensor malfunction 23 mA
Under range dew point 4 mA
Exceeding the dew point range by 20 mA
Approved electric isolators KFD2-CR-EX1.20200
KFD2-CR-EX1.30200
KFD0-CS-EX2.50P
KFD0-CS-EX1.50P
KFD2-STC4-EX1.H
MTL5040
MTL5541