MKS Baratron ® Single ended pressure switch(41C)
The 41C single ended 10-26000 Torr pressure switch provides accurate and reliable protection for vacuum/pressure processes. These switches are designed for applications that do not require DC signal output, providing relay outputs that can be easily connected to alarms, valve actuators, computers, process controllers, load locks, and other protective devices. 41C is the reference atmospheric pressure. The trigger point can be set to be above, below (reverse calibration), or just triggered at the current atmospheric pressure.
lSingle end configuration with reference to atmospheric pressure
lExcellent set point accuracy and low lag
lThe all metal and all welded structure only uses stainless steel andInconel®To resist corrosion
lThe switch relay can be set to be energized above or below the set point to achieve fault safe operation
lHigh reliability reduces downtime
MKS Baratron ® Single ended pressure switch(41C)
Technical Specifications
type |
pressure switch |
temperature coefficient |
Full range ±0.07%/°C |
configuration |
single-ended |
Materials exposed to process gases |
Chromium nickel iron alloy and316Lstainless steel |
accuracy |
Full range±0.5%(±Temperature coefficient |
overpressure |
2 xFull scale or45 psiaThe larger one shall prevail |
Jump point range |
Full range5%to100% |
output |
Electromechanical relay- SPDT(Isolation) contacts,30 VDCThe rated current of the resistor is as high as1 Amp |
Jump point dead zone |
Full range ±3% |
Power requirements |
20-30Volt direct current |
response time |
<20millisecond |
Working environment temperature |
0-50℃ |
Internal volume |
3.3cc |
comply with |
CETheULrecognize* (cULus E520864)TheEN/IEC CBApprove all country deviations*Standard onlyMKSPart Number |
feature
Baratron ® Capacitive pressure gauge technology
Capacitive pressure gauge is an electromechanical pressure gauge that can measure pressure and vacuum. The capacitance meter converts the pressure modulated motion in the thin diaphragm into an electrical signal proportional to the pressure. A pressure sensor is a thin diaphragm exposed to pressure or vacuum measured through an inlet tube. The electrode is installed in the reference chamber behind the diaphragm. The pressure difference between the process and the reference chamber causes the diaphragm to slightly deflect, changing the distance between the diaphragm and the electrode. The change in distance will cause a change in capacitance between the diaphragm and the electrode, resulting in an electrical signal proportional to the pressure change. Due to the fact that the difference in capacitance signals is caused by physical changes within the pressure gauge, rather than changes in gas characteristics, the pressure measurement of the capacitance pressure gauge is independent of the composition of the measured gas.

Reference atmospheric pressure
41 and42 vacuum/The pressure switch refers to atmospheric pressure. They are usually used to ensure that the load lock pressure is balanced with the local atmospheric pressure before opening the door.MKSThe trigger point on the atmospheric switch can be set to be above, below (reverse calibration), or just triggered at the current atmospheric pressure.
Factory settings jump point
The relay trip point is set to the full range pressure range at the factory5%to100%This means that personnel do not need to adjust the set points, and there will be no safety issues caused by incorrectly adjusted set points.
Fault safety relay operation
41、42、51 and52 The relay mode on the switch can be set to power on when it is above or below the set point. If the device loses power, the relay will switch to the normally closed position. Users can indicate whether the normally closed position is above or below the set point. Taking electricity above the set point as an example, when the pressure is higher than the trip point, the relay is in the normally open position, while when the pressure is lower than the trip point, the relay is in the normally closed position. For“Electricity below the set point”Option, the situation is exactly the opposite. In a vacuum system, fail safe operation refers to the situation where if the system loses power and the relay loses power, the relay will be in the same state as the high voltage state. Therefore, most vacuum systems require relays to be energized when the pressure drops or falls below the set point.
41BThe42BThe51Band52BVacuum and atmospheric pressure switches provide higher precision than mechanical switches, thereby providing stricter process control and repeatability, thereby improving throughput and yield.
9 needle“D”Type pin arrangement
safety pin |
description |
1 |
Power return(-) |
2 |
Power input(+) |
3 |
Relay normally open contact |
4 |
Relay common terminal |
5 |
Relay normally closed contact |
6 |
Unused |
7 |
Unused |
8 |
Unused |
9 |
Chassis grounding |
Flywire wiring
wire |
description |
red |
Power input(+) |
black |
Power return(-) |
green |
Relay normally open contact |
white |
Relay common terminal |
orange |
Relay normally closed contact |
bare wire |
Chassis grounding |
Hangzhou Yushi Technology Co., LtdFocusing on providing technical services and solutions related to flow, pressure, vacuum detection and control for industrial and scientific research customers, the company acts as an agent and distributor for brands such as ALICAT in the United States, Vogtlin in Switzerland, MKS in the United States, and EBARA in Japan. Combining the products we represent, we provide customers with high-quality flow and pressure monitoring solutions, aiming to improve their research and production efficiency, enhance their manufacturing processes, and promote their scientific research and innovation progress.