VLG10 is a liquid vaporization device that can vaporize liquid samples (such as fragrances, aromatic agents, odor standards, etc.) into gas phase samples. During the process of liquid vaporization, the amount of liquid, vaporization temperature, vaporization time, and gas volume flow rate can all be quantitatively controlled to obtain standardized gas samples for subsequent analysis, such as odor concentration detection, odor intensity and pleasure evaluation, sample composition analysis, etc.
1. Preparation of high boiling point standard samples and tubes;
2. Determination of perception threshold;
3. Comparative analysis of odor samples;
4. Determination of odor sample components;
5. Optimize sample formula, etc.
| PRODUCT MODEL | VLG10 |
| ★Vaporization temperature | Temperature range: Room temperature+10~250 ℃ |
| ★Vaporization flow rate | 0~5000mL/min |
| ★Flow accuracy | ±1%FS |
| Vaporization volume | 0~999L |
| Mpa | 0.2~0.4MPa |
| Working power supply | AC220V±10%,10A,50/60Hz |
| Liquid inlet method | Manual injection (Figure 1)![]() Figure 1 Manual liquid inlet |
| ★Export method |
Polytetrafluoroethylene two-way connector: diameter 6.35mm (other specifications can be customized)
It can be connected to sampling bags (Figure 2) and sampling tubes (Figure 3) for gas collection, and can also be connected to sniffing ports (Figure 4) for odor evaluation ![]() Figure 2 Sampling bag Figure 3 Sampling tube
![]() Figure 4 Smell and distinguish mouth
|
| Product size | 260*360*160mm |
| Product weight | 6KG |