Labom pressure transmitter - model: COMPACT, imported

NegotiableUpdate on 12/12
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
Compact pressure transmitter COMPACT is a high-precision converter that converts pressure measurement values into load independent current signals (e.g. 4... 20 mA). The model series CC60. - F is designed to meet the strict requirements of the food processing, pharmaceutical technology, and biotechnology industries. Labom pressure transmitter - model: COMPACT, imported
Product Details

Labom pressure transmitter - model: COMPACT, imported

Labom pressure transmitter - COMPACTDesign features

  • Range 0... 100mbar to 0... 100bar
  • Linear error lag <+0.2% f.s
  • Varistor measurement system
  • Sanitary design complies with EHEDG, FDA, and GMP regulations
  • The material and surface quality meet the hygiene requirements
  • The liquid receiving component is made of stainless steel and welded
  • Standard stainless steel casing or outdoor casing
  • Protection level IP65, optional IP67
  • Signal output 4... 20mA, optional 0... 20mA
  • Optional when the process temperature reaches 200 ℃
  • Gas explosion-proof protection
  • Compliant with SIL2 level
  • Inspection certificate: Material certificate complies with DIN EN 10204-3.1

Labom pressure transmitter - COMPACTTechnical data

Shell design

design

Outdoor housing IP65 or IP67 with cable connectors

A right angle plug that complies with DIN EN 175301-803-A (DIN 43650, Form A), IP65

Cable connection, IP67

Circular connector M12, IP65

Connecting nut: Polyamide (with electrical plug or cable connection) with silicon seal for electrical components

The temperature impact within the relevant temperature range of 0... 50 ℃:

a) Shell - Zero point<0.2%/10 K fs

-Distance measurement<0.2%/10 K f.s

b) Process connection (diaphragm seal) according to design

Zero point error of flat packaging film box

DN25/ 1″ 4.8 mbar/10 K

DN32/ 1 1/2″ 2.3 mbar/10 K

DN40 1.6 mbar/10 K

DN50/ 2″ 0.6 mbar/10 K

Online membrane box zero point error

DN25/ 1″ 9.5 mbar/10 K

DN32/ 1 1/2″ 4.1 mbar/10 K

DN40 3.9 mbar/10 K

DN50/ 2″ 3.9 mbar/10 K

The zero point error of the listed process connections is a reference value for standard design. We can provide a detailed system calculation as requested. We can also provide diaphragm sealing systems with small errors.Labom pressure transmitter - model: COMPACT, imported

Auxiliary power supply

Standard design of pressure transmitter

·Nominal voltage 24V DC

Internal cavity ventilation is used for ranges<16 bar through shell threads or connecting cables (according to design)

Shell material: stainless steel

Process Connection

See the parameter values on the next page or order code

Material of casing: No. 1.4404 (316L)

Temperature range

Environmental temperature: -25... 70 ℃

Storage temperature: -40... 90 ℃

·Functional range 6... 30V DC

·Maximum allowable voltage 30V DC

Influence of power supply voltage

≤0.01% f.s. / V

Signal output

4…20mA, 2-wire circuit

Current limit of output signal

The maximum output current is about 30mA

Adjustment scope

Approximately ± 5% f.s. Zero point and distance measurement can be adjusted separately

Load

Standard design

UB=working voltage

Ra=maximum allowable load resistance (including wires)

Load impact

500 ohm load variation: ≤ 0.1% f. s

Safe operation

EN 61508, level compliant with SIL 2, T Ü V-Reg. - No. 44 207 1038 1144

Process temperature: see selection table

Range/Overload Limit

The intermediate range can be adjusted as required

Response time

≤20 ms

Measurement accuracy of pressure transmitter

Linear error includes hysteresis<0.2% f.s

(For ranges ≥ 0... 60bar, it is less than 0.3% f.s.)

Fixed point adjustment, with an adjustment accuracy of<± 0.2% fs

Explosion proof certification

CENELEC certification complies with ATEX Explosion proof Intrinsic Safety Government Testing Laboratory (T Ü V) 00ATEX 1557 X

. II 2G EEx ib IIC T6

. Umax ≤ 30V DC

. Imax ≤ 150 mA

. Pmax ≤ 1 W

. Ci ≤ 49 nF

Band Li ≤ 33 μH

Weight (without diaphragm seal)

Outdoor shell: approximately 460 g

Shell with connecting plug: approximately 200 g

Installation location

arbitrary

EMC testing

Noise interference complies with EN50082 Chapter 2, March 1995 edition, industrial standard

Radiation interference complies with EN50081 Chapter 1, 1993, residential and industrial area standards

The launching device itself is non radioactive

Keywords