L180 Coating Thickness Gauge (Eddy Current Method)

NegotiableUpdate on 03/15
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
The L180 coating thickness gauge (eddy current method)/oxide film thickness gauge produced by LEE SYSTEMS in the United States adopts world advanced technology and is an ultra small coating thickness gauge. The L180 coating thickness gauge (eddy current method) can non destructively measure the thickness of non-conductive coatings on non-magnetic metal substrates (such as copper, aluminum, zinc, tin, etc.) (such as rubber, paint, plastic, anodic oxide film, etc.). The L180 coating thickness gauge (eddy current method) has a one click operation function, which does not require calibration or resetting to perform the next measurement, and does not require training for operators to operate.
Product Details
L180 Coating Thickness Gauge (Eddy Current Method)

The L180 coating thickness gauge (eddy current method) can non destructively measure the thickness of non-conductive coatings (such as enamel, rubber, paint, plastic, etc.) on non-magnetic metal substrates (such as copper, aluminum, zinc, tin, etc.).

The L180 coating thickness gauge (eddy current method) can be widely used in manufacturing, metal processing, chemical industry, commercial inspection and other testing fields. Due to its small size and integrated probe, this instrument is particularly suitable for on-site engineering measurements; This instrument is sturdy, durable, versatile, and can be operated with one hand.

L180 Coating Thickness Gauge (Eddy Current Method)
Designed to be portable and durable, enabling it to function properly under harsh conditions. It has a one click operation function, which does not require calibration or resetting to perform the next measurement, and does not require training for operators to operate.


Function:

Zero calibration and full range calibration can be performed.

Ruby anti-wear probe.
Basic calibration can be performed on the probe.
It has an automatic shutdown function.
There is a buzzing sound during the operation process.
There is an undervoltage indication function.
There is an error prompt function.


Technical Specifications:

Measurement principle: Eddy current method
Measurement range:11250μm
Measurement accuracy:±[(13)%H+1]μm(Full range calibration)

Resolution:1μm

Workpiece requirement: * small curvature radius(mm)Convex5mm/concave10mm

Contact * small area diameter10(mm)

Thickness of the substrate facing the street(mm 0.05

Usage environment: Temperature0-40centigrade;

humidity20%-75%; No magnetic field environment

Power supply: four sections1.5ambushAAAbattery

Overall dimensions:112 ×69×28mm

Weight:82 g(Battery not included)


Standard configuration:

Host, standard calibration sheet set, calibration aluminum substrate, usage guide, English inspection report certificate, suitcase, lanyard, and battery, etc.


Optional attachments: