KEW 2433 leakage current clamp meter | KEW 2433 leakage current tester

NegotiableUpdate on 03/25
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
2. RMS true RMS test (2433R).
Product Details

The KEW 2433 leakage current clamp meter has the following advantages:

1. Advanced performance design with minimal external magnetic field impact.

2. RMS true RMS test (2433R).

3. When exchanging about 10mA, a conductor with a diameter close to 15mm carries an alternating current of 100A.

4. Three AC current ranges: 40mA/400mA/400A.

5KEW 2433 leakage current clamp meter data retention function.

6. The peak holding function and automatic shutdown function save battery.

KEW 2433 leakage current clamp meter | Technical specifications of KEW 2433 leakage current tester:

This parameter table is provided byKyoritsuFlukeTESTODeltaTRAKDistrict General AgentShenzhen Haixu Instrument and Meter Co., Ltd. provides original and authentic products

Measurement range

accuracy
AC current (50/60Hz)
40/400mA/400A
± 1% ± 5 digits of reading (40/400mA)
± 1% ± 5 digits of the reading (0-300A)
± 2% of the reading (300.1~399.9A)
Alternating Current (WIDE)
40/400mA/400A
± 2.5% ± 10 digits of reading [20-1KHz], ± 1% ± 5 digits of reading [50/6OHz] (40/400mA),
± 2.5% ± 10 digits of reading [40-1KHz], ± 1% ± 5 digits of reading [50/60Hz] (0-350A),
± 5% of reading [40-1kHz], ± 2% of reading [50/60Hz] (350.1-399.9A)
frequency response
20Hz~1KHz(40Hz~1KHz:400A)
Maximum circuit voltage
AC/DC 600V (between the measuring line and the neutral line), AC/DC 300V (grounded)
Conductor diameter
Maximum diameter of 40mm
Safety specifications
IEC61010-1 CAT. III 300V Pollution Level 2 IEC61010-2-032
External magnetic field influence
When exchanging about 10mA, the AC current carried by a conductor with a diameter close to 15mm is 100A
withstanding voltage
3700V AC/min
response time
About 2 seconds
power supply
RO3 or equivalent (1.5V DC) battery x 2
External dimensions
185(L)×81(W)×32(D)mm
weight
About 270g
attachment
9052 (Carrying Case) R03 (1.5V) × 2 User Manual
option
8004/8008 (Converter - cannot be used for leakage current testing)

Instrument recommendation: MG125 insulation resistance tester DG6 insulation resistance tester