Jenco 9173R desktop dissolved oxygen analyzer

NegotiableUpdate on 05/14
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

Jenco 9173R desktop dissolved oxygen analyzer

Product Details

Product Introduction

Product Introduction

9173R dissolved oxygen, temperature, desktop tester
  • 9173R can measure dissolved oxygen, air saturation, and temperature
  • Microcomputer processor with high measurement accuracy and good repeatability
  • Ultra large LCD display, temperature display, low battery warning
  • AC/DC dual-use, universal electrode bracket, easy and simple operation
  • RS-232C communication interface
  • Imported polarographic dissolved oxygen electrode

specifications

  • display range resolution accuracy
    dissolved oxygen 0.00 to 20.00 ppm 0.01 ±0.2%FS
    air saturation 0.0~200.0% 0.1 ±0.2%FS
    temperature -6.0~46.0℃ 0.1 °C ±0.3 °C
  • type of electrodes Polarographic electrode
    Dissolved oxygen temperature compensation range Automatic -6.0~46.0 ℃
    Pressure Compensation 600~1100mBar adjustable, default value is 1013mBar
    Salinity compensation 0.0~40.0pppt adjustable, default value is 0.0ppt
    Correction data storage have
    communication RS-232C
    Sound feedback All buttons
    power supply AC power converter (output: DC9V) or 6 1.5 volt 5 # batteries
    work environment 0~50 ℃, relative humidity<>
    size 230x165x80mm
    weight About 650g (excluding battery)
    Standard configuration
    9173R Host, LD-900-7 electrode, LD-900-3A electrode membrane head, 007N universal bracket
    AC power converter (output: DC9V), 6 1.5 volt 5 # batteries, software CD