JB3126A Intelligent Insulation Resistance Tester
JB3126A Intelligent Insulation Resistance TesterisHigh voltage insulation resistance meter with a measuring range of 5 for measuring insulation resistance.
● Design in compliance with international safety standards:
IEC 61010-1, -2-030 (CAT III 600V/CAT IV 300V Pollution Level 2)
IEC 61010-031 (Requirements for Handheld Probes)
● Automatic discharge function: When testing the insulation resistance of capacitive loads, the charging charge is automatically released after measurement. The discharge state can be confirmed in voltage mode.
● The backlight function is convenient for working in low light or at night.
● A bar chart displays the measurement results.
● Warning signs and buzzing warnings for live lines.
● Automatic shutdown function: To avoid battery waste caused by forgetting to shut down, after measurementThe instrument will automatically shut down when there is no operation for 10 minutes.
● Automatic measurement andPI measurement (polarization index measurement), DAR measurement (induced absorption ratio measurement).
3.Technical Specifications
Measurement range and accuracy(temperature/humidity: 23±5ºC,45~75%RH) <Insulation resistance test>:
rated voltage |
250V |
500V |
1000V |
2500V |
5000V |
measurement range |
0.5MΩ~100.0MΩ |
0.5MΩ~ 99.9MΩ 80~1000MΩ |
0.5MΩ~99.9MΩ 80~999MΩ 0.80~2000MΩ |
0.5MΩ~99.9MΩ 80~999MΩ 0.80~10000MΩ 8.0~100000MΩ |
0.0005GΩ~0.1GΩ 0.08~1GΩ 0.80~9.99GΩ 8.0~99.9GΩ 80~2000GΩ |
Display range |
0.5MΩ~100.0 MΩ |
0.5MΩ~1000 MΩ |
0.5MΩ~2000 MΩ |
0.5MΩ~100000 MΩ |
0.0005GΩ~2000 GΩ |
open-circuit voltage |
DC250V +10% ,-10% |
DC 500V + 20%,-10% |
DC 1000V + 20%,-0% |
DC 2500V +20%,-0% |
DC5000V +20%,-0% |
|
Freeze measurement of current
|
At a load of 1M Ω 1mA~1.2mA |
At a load of 2.5M Ω 1mA~1.2mA |
At a load of 5M Ω 1mA~1.2mA |
||
short-circuit current |
5mA±0.5mA |
||||
accuracy |
±5%±3dgt |
0.5MΩ~200GΩ:±5%±3dgt Greater than 200G Ω~20000G Ω:±20%
|
|||
Voltage monitor (insulation resistance range)
30~600V(resolution10V): ±10%rdg±20V
This mode is suitable for confirming the discharge state of the charged charge in the tested object. When measuring, use the measured voltage value displayed on the monitor as the standard. Please pay attention to the external application of AC voltage
The displayed data is not the correct value.
<voltage test>
DC voltage |
AC voltage |
||
measurement range |
±30 -- ±600V |
30 -- 600V(50/60Hz) |
|
resolution |
1V |
||
accuracy |
±2%rdg±3dgt |
||
display: |
LCD (Liquid Crystal Display) Insulation resistance range: up to 1200 counts VAC/DC range: up to 630 counts Bar chart: maximum 36 points DAR/PI value: maximum 9.99 Time: Maximum 99.59 |
Low battery warning: |
Battery logo (4 levels) |
|
Over range display:
|
Display the "OL" symbol (insulation resistance range) Display the "Hi" symbol (voltage range) |
|
Automatic range:
|
Move to higher range: 1000 counts Move to lower range: 80 counts (insulation impedance range only) |
Automatic shutdown function: |
Automatic shutdown after 10 minutes of operation (this function does not work during testing) |
Environmental conditions for use: |
Below an altitude of 2000m |
Accuracy guarantee temperature and humidity range: |
23 º C ± 5 º C/Relative humidity below 85% (no condensation) |
Operating temperature and humidity range: |
0 º C~40 º C/relative humidity below 85% (no condensation) |
Storage temperature and humidity range: |
-20 º C~+60 º C/relative humidity below 75% (no condensation) |
Overload protection: Insulation resistance range: |
AC1200V/10 seconds. |
Voltage endurance: |
Voltage range: AC720V/10 seconds. AC5160V (50/60Hz)/5s (between circuit and outer box) |
Insulation impedance: |
≥ 1000M Ω/DC 1000V (between circuit and outer box) |
Size: |
177(L)×226(W)×100(D)mm |
Weight: |
1.9kg (including battery) |
Power Supply: |
DC12V: alkaline battery size C (LR14) x 8 cells |
● current consumption:(Power supply voltage:12VThe representative value of time
range |
250V |
500V |
1000V |
2500V |
5000V |
VAC/DC |
When there is an output short circuit |
600mA |
110mA |
||||
At rated current output |
650mA/0.25MΩ |
800mA/0.5MΩ |
1000mA/1MΩ |
1350mA/2.5 MΩ |
1350mA/5 MΩ |
|
When the output is open |
40mA |
40mA |
50mA |
80mA |
150mA |
|
standby |
25mA |
|||||
When the backlight is on |
Increase 40mA |
|||||
●measure time: about25hours
insulation resistance 5000V range 100 MΩ time