- Phone
-
Address
508B, Building D, Xinyuan Science and Technology Park, No. 97 Changping Road, Changping District, Beijing
Beijing Spectrum Century Technology Development Co., Ltd
508B, Building D, Xinyuan Science and Technology Park, No. 97 Changping Road, Changping District, Beijing
WS-5937-JZ2Aintermediate frequencySensor calibration system |
1.System Introduction and Application
When designing or using sensors, the frequency response characteristics and sensitivity of the sensor are quite important technical indicators. This product is designed specifically for calibrating the frequency response characteristic curve and sensitivity of speed and acceleration sensors. With this system, automatic closed-loop control of the "sensor frequency response characteristic curve calibration" can be achieved by computer, measuring the frequency amplitude of standard sensors and calibrated sensors, calculating the frequency response and sensitivity of calibrated sensors, and completing the calibration of the frequency response characteristic curve according to user requirements.
This system can be applied to metrology research institutes, vibration research laboratories, sensor research and production units, etc.
By utilizing proprietary technology, this system can also calibrate the sensitivity and frequency response curve of speed sensors (detectors), which are widely used in fields such as petroleum geophysical exploration and mechanical vibration monitoring in factories and mines.
Due to our configurationWS-5937The control frequency of the calibration controller can reach50kHz,But currently there is no suitable high-frequency vibration table, so we only achieve intermediate frequency calibration. If the user has their own high-frequency vibration table, they can also choose our calibration system.

2.System Composition
2.1 WS-5937Calibration controller1Taiwan;
2.2 WS-PA-30power amplifier1Taiwan;
2.3 Intermediate frequency exciterJZ-2AorJZ-20C 1Platform (optional);
2.4 Sensor calibration softwareVib’SIN 1Set;
2.5 Standard sensors1Only;
2.6 computer1Taiwan (self provided);
3.software interface

4.System composition logic diagram

5.function
5.1 Scan frequency points and amplitudes according to user customization;
5.2 The software automatically generates scanning curves, with equal frequency intervals and equal amplitude scans;
5.3 Automatic control of amplitude error;
5.4 Automatically generate frequency response characteristic curve data (volt frequency, phase frequency);
5.5 Frequency response characteristic curve plotting (linear, logarithmic);
5.6 Programmable signal generator, waveform selection: sine wave, single frequency signal output, signal sweep, frequency dwell;
5.7 Online assistance;
6.technical indicators
6.1 WS-5937Calibration controller
6.1.1Data collection and control signals:
1)ComputerUSBInterface control and collection;
2)Data collection frequency200kHz;
3)16bit4Road data collection, input range ±10V;
4)1road16bitD/ADigital to analog conversion signal output channel;
5)Control signal output amplitude ±10V, amplitude resolution16Position, amplitude control accuracy0.1%FS;
6)Control the sweep frequency range1Hz~20kHz, frequency resolution28Position, frequency control accuracy0. 1Hz;
7)Supply voltage:100V~240V(AC);
8)Size: Width365X high70X deep365mm;

6.1.2Charge amplifier:
1) OLEDDisplay parameters;
2)One encoder key sets all charge amplifier parameters;
3)Number of channels:2通道;
4)Input range: ±100000pC;
5)Gain:5档;
6)Normalization:0.1~999;
7)Accuracy: Acceleration:≤±0.5%;
speed:≤±0.5%;
displacement:≤±2.0%;
8)Frequency range: Acceleration: 0.3Hz~50kHz;
speed: 1.0Hz~10kHz;
displacement: 1.0Hz~1kHz;
9)Low pass filter: ① Filtering method: Programmable8Order continuous filtering;
② Programmable low-pass filtering frequency:1Hz~50kHz(arbitrarily set)
③ Filter stopband attenuation:-80dB/Oct;
10)Output voltage: ±10V
11)Overload indication: Input voltage exceeds ±9.8VtimeLEDThe indicator light is on;
12)Output noise: ≤5×10-3pC(Maximum gain converted to input terminal)
13)Input impedance:1012Ω
14)Output resistance:<1Ω
15)Output current:15mA
16)Working temperature and humidity:-10~50℃、<85%RH(No coagulation)
6.1.3 IEPEadapter
1 ) OLEDDisplay parameters;
2 )One encoder key to set allIEPEAmplifier parameters;
3 )Number of channels:2通道;
4)Voltage input range: ±5V;
5)Gain:5档;
6)Accuracy: Acceleration:≤±0.5%,
speed:≤±0.5%,
displacement:≤±2;
7)Frequency range: Acceleration: 0.1Hz~50kHz,
speed1Hz~10kHz,
displacement: 1Hz1kHz;
8)Low pass filter: filtering method: programmable8Order continuous filtering,
Programmable low-pass filtering frequency:1Hz50kHz(arbitrarily set),
Filter stopband attenuation:-80dB/Oct;
9)Overload indication: Input voltage exceeds ±9.8VtimeLEDThe indicator light is on;
10)Signal to noise ratio:>50dB;
11)Noise ≤5mVrms;
13)Output resistance: <1Oh?
14)Output current:15mA;
15)Working temperature:-10~50℃;
6.2 WS-PA-30power amplifier
1)Output power:30W;
2)Maximum output current:3A;
3)Maximum output voltage: ±10V;
4)Frequency range:10~20kHz;
5)Output impedance:1Oh?
6)Distortion degree:<2%;
7)noise<5mV;
6.3Intermediate frequency exciter
6.2.1 JZ-2ABasic technical specifications of exciter(choice1)
1)Working frequency:30~10KHz;
2)Current limit:2.5A;
3)Carrying He:100g;
6.2.2 JZ-20CBasic technical specifications of exciter (selection)2)
1)Working frequency:10~2000Hz;
2)Maximum displacement:±10mm;
3)Maximum current:10A;
4)He Carrying:1000g
6.4Standard accelerometer indicators (optional)
6.3.1Model:BK-8305Optional
Manufacturer: DenmarkB&K;
Compliant with standards:ISO 5347;
Sensitivity:1.25pC/g;
Working frequency:1~5000Hz;
Resonant frequency:35kHz;
Maximum measurement:1000g;
Weight:45g;
Operating Temperature-74℃~250℃;
Application: Used for sensitivity calibration of sensors;
Explanation: Due toBK-8305The price is relatively high, you can choose other brands or domestic acceleration sensors with appropriate prices, and then regularly calibrate them at the measurement and testing unit for sensor calibration.
6.3.2Model:CK-8305Optional
Manufacturer: Beijing Spectrum Company;
Sensitivity:1.25pC/g;
Working frequency:1~5000Hz;
Resonant frequency:35kHz;
Maximum measurement:1000g;
heavy Quantity:45g;
Operating Temperature-20℃~80℃;
should Used for sensitivity calibration of sensors
7.Order model and quotation
serial number |
name |
model |
cycle |
remark |
1 |
Intermediate frequency sensor calibration system(30~10KHz) |
WS-5937-JZ2A |
2-4week |
Standard configuration, including1Only standard acceleration sensorCK8305 |
2 |
Intermediate frequency sensor calibration system(10~2000Hz) |
WS-5937-JZ20C |
2-4week |
Standard configuration, including1Only standard acceleration sensorCK8305 |
8.Intermediate frequency vibration tableJZ-2ASelf frequency response curve

Intermediate frequency vibration tableJZ-2Afrequency response curve(30~11000Hz)
9.Example of frequency response characteristic curve of sensor
9.1 example1:Frequency response and phase curve of a certain accelerometer

Frequency response and phase curve of a certain accelerometer30-10000Hz
9.2 Example 2:Frequency response and phase curve of a certain accelerometer

Frequency response and phase curve of a certain accelerometer10-2000Hz
10.Example of frequency response characteristic curve of intermediate frequency calibration sensor

Frequency response calibration curve of acceleration sensor(Calibrate frequency response100~5000Hz)Measure its resonance peak at3.5kHz
11.Calibration question: Is the system accurate?
Regardless of whether the power amplifier and exciter used in the calibration system are domestic or imported instruments and equipment, the system always has nonlinearity. However, the sensor calibration adopts the comparison method (usually referred to as the "back-to-back method"), and the calibration is based on the standard sensor to calibrate other sensors. This system recommends using Denmark as the referenceB&Kcompany2Only standard sensors,B&K-4374SMainly used for calibrating the frequency response curve of sensors, andB&K-8305Mainly used for calibrating the sensitivity of sensors and the frequency response curve of low-frequency sensors, currently the domestic metrology department mainly adopts DenmarkB&K-8305As a standard sensor, DenmarkB&K-8305The sensor adoptsLaser absolute method calibration meets international standardsISO 16063-11 (previousISO 5347)The requirements.

12.Triple spectrum of acceleration, velocity, and displacement
They can be identified from the triple spectrum of acceleration, velocity, and displacement3A value at a certain frequency(Hz)The corresponding relationship below. From the figure, it can be seen that at low frequencies, the displacement is relatively large; At high frequencies, the displacement is very small.
Unit and color in the figure: orange accelerationAcceleration(m/s2), Black SpeedVelocity(mm/s)Blue displacementDisplacement(mm)Black frequencyFrequency(Hz).