Intelligent softening breakdown testerProduct performance characteristics:
1. Implement GB/T4074.6/IEC 60851-6 standards and NEMA and JIS standards in the United States and Japan
2. Inspection standard: JB/T4279.8
3. Suitable for testing the insulation performance of enameled round wires and thin film wrapped round wires with a nominal diameter of 0.020mm to 3.000mm and below, after softening in a hot state
4. There are three testing methods: constant temperature method, temperature rise method 1, and temperature rise method 2
5. Automatic temperature measurement, control, adjustment, and display; According to the set preheating time, target temperature, and test time, the instrument automatically completes the preheating, heating, load application, and test voltage of the sample. When the test time is reached or the sample is broken down, the load is automatically unloaded and an alarm is triggered, and the test time and temperature are displayed. The entire testing process is automatically completed
6. Automatic forced air cooling fan enables rapid second test
7. The heating mold is made of brass material
Weights equipped:
Weight holder 0.25N × 1, 1N × 1
砝码0.15 N×1、0.25N×1、0.45N×1、1.2N×1、3.5N×1、4.5N×2、17N×1、34N×1、35N×1
Optional models:
RJ-20 softening breakdown tester
Intelligent softening breakdown testerMain technical parameters:
Heating mold size |
100×100×70mm |
Heating mold, intersection angle of two wire slots |
90° |
Test voltage between two samples |
Constant temperature method and rising temperature method 1: AC100 ± 10V |
Heating method 2: AC155V ± 15V | |
Sample breakdown action current |
Fixed temperature method and rising temperature method 1:5 ± 0.5mA |
Heating method 2:20 ± 10mA | |
Short circuit current between samples |
<50mA |
Display temperature and temperature difference at the intersection of the sample |
<2℃ |
Sample preheating time, test time |
1 second to 9999 seconds can be set arbitrarily |
Max test temperature |
≤500℃ |
Temperature setting for constant temperature method test |
RT~500 ℃ can be set arbitrarily |
Error of timing device for 120 seconds |
±1s |
input power |
AC220V±10% 50Hz |
heating power |
900W |
Overall dimensions |
L×W×H 540×310×345mm |
weight |
32kg |