Welcome Customer !

Membership

Help

Shenzhen Shangyoutai Technology Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Products

Shenzhen Shangyoutai Technology Co., Ltd

  • E-mail

  • Phone

  • Address

    9th Floor, Building 5, Dongfang Community, Songgang Street, Bao'an District, Shenzhen

Contact Now

Intelligent PH/ORP/Temperature Three in One Analyzer

NegotiableUpdate on 06/24
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

Intelligent three in one analyzer, also known as industrial pH/ORP/temperature online analyzer. It is a brand new pH, ORP, and temperature analyzer with high intelligence and flexibility. It can simultaneously measure pH, ORP, and temperature, and is widely used in urban sewage treatment plants, power, water supply, pharmaceuticals, chemical, food and other industries to continuously measure solution pH, ORP values, and temperature values.

Product Details

Intelligent three in one analyzerPC-3218 Industrial PHORP, temperatureOnline analyzer is a brand new pH analyzer/ORP/TemperatureThe analyzer has a high degree of intelligence and flexibility, and can simultaneously measure pH/ORPAnd temperature, widely used in urban sewage treatment plants, electricity, water supply, medicine, chemical, food and other industries, to regulate the pH of solutions/ORPContinuous measurement of values

Basic functions

1.The factory standard interface in both Chinese and English allows for easy switching and simple operation.

2.Automatic and manual temperature compensation can be selected to meet various user needs.

3.Two 4-20mA outputs, corresponding to PH &ORPValue, using isolation technology, strong anti-interference ability.

4.The high and low points of the two relays can be switched freely, and the hysteresis can be adjusted freely.

5. 480*800 IPSLCD, clear display,10-100%Brightness can be freely set.

6.Protection level IP65, suitable for outdoor use.

7.Password management function to prevent misoperation by non professionals.

Instrument technical parameters

Measurement range: PH (0-14PH) ORP (-1999-+1999 MV).

Accuracy:+0.01 pH;+1 mV.

Resolution: 0.01pH; 1mV.

Stability: ≤ 0.02 pH/24 hours; ≤ 3 mV/24 hours.

PHStandard solution: 4.01-6.86-9.18, 4.00-7.00-10.00, two-point or three-point calibration.

Temperature compensation: 0-100 ℃ manual/automatic (PT1000).

Signal output: 4-20mA isolated protection output, independently corresponding to PH &ORPThe maximum load is 300 Ω.

Alarm output: Two groups can correspond to high and low point alarms (3A/250 V AC) at will, with normally open contact relays.

Power supply: AC 220V.

Power consumption: ≤ 5W.

Environmental conditions: (1) Temperature 0-60 ℃ (2) Humidity ≤ 85% RH.

Dimensions: 144 × 144 × 115mm (height × width × depth).

Hole size: 138 × 138mm (height × width).

Protection level: IP65