This product uses the Android operating system and is matched with a 10.1-inch capacitive touch screen, with simple and smooth operation; Forward and backward air ducts, machines can be placed side by side; Adaptive pressure rod hot cover, one step closing and tightening the cover; Adopting a long-life Peltier module specifically designed for temperature cyclers, with a maximum temperature change rate of 6 ℃/S; equipped with a built-in WIFI module, multiple machines can be controlled through a mobile app.
RePure-EIntelligent Gradient Gene Amplification DeviceAdopting the Android operating system, matched with a 10.1-inch capacitive touch screen, the operation is simple and smooth; Forward and backward air ducts, machines can be placed side by side; Adaptive pressure rod hot cover, one step closing and tightening the cover; Adopting a long-life Peltier module specifically designed for temperature cyclers, with a maximum temperature change rate of 6 ℃/S; equipped with a built-in WIFI module, multiple machines can be controlled through a mobile app.
Instrument Features
1). Adopting a long-life Peltier module specifically designed for temperature cyclers;
2). The engineering reinforced aluminum module treated with anodizing technology retains both rapid thermal conductivity and sufficient corrosion resistance;
3). Fast heating and cooling rate, with a maximum heating and cooling rate of 6 ℃/s, saving valuable experimental time;
4). Adaptive pressure rod hot cover, one step closing and tightening of the cover, can adapt to test tubes of different heights;
5). Forward and backward air ducts, machines can be placed side by side;
6). Adopting the Android operating system, matched with a 10.1-inch capacitive touch screen, a graphical menu style navigation interface, and extremely simple operation;
7). Built in 11 standard program file templates for quick editing of required files;
8). Users can customize 9 shortcut buttons on the homepage according to their usage needs;
9). Real time display of program progress and remaining time, supporting intermediate programming during PCR instrument operation;
10). One click rapid incubation function, meeting the needs of denaturation, enzyme digestion/ligation, ELISA and other experiments;
11). The temperature and working mode of the hot cover can be set to meet different experimental needs;
12). Automatic power-off protection, automatically executes incomplete cycles after power restoration, ensuring safe operation of the entire amplification process;
13). USB interface, supports USB inventory/retrieval of PCR data, and can also control the PCR instrument using a USB mouse;
14). Support USB and LAN software updates;
15). Built in WIFI module, allowing one computer or mobile phone to control multiple PCR machines simultaneously through network connection;
16). Support the function of sending email reminders when the experimental program ends.
performance metrics
| Product Model |
RePure-E |
| sample size |
96×0.2ml
|
| test tube |
0.2mlSingle tube,8Combo pipe, 12 Combo pipes, 96 hole microplate, compatible with no skirt edge and half skirt edge96KongPCR reaction
|
| reaction system |
5-100ul |
| temperature range |
0-105℃
|
|
maximumliftTemperature rate
|
6℃/s
|
| temperature uniformity |
≤±0.2℃
|
| Temperature accuracy |
≤±0.1℃
|
| Temperature display resolution |
0.1℃
|
| Temperature control method |
Block\\Tube |
| Adjustable temperature change rate |
0.01-6℃/s
|
| Gradient temperature range |
30-105℃
|
| Gradient type |
Conventional/Linear |
| Gradient setting range |
horizontal12Column: 01-42℃ |
| Hot cover temperature range |
30-115℃
|
| Number of program storage |
200000+(USB FLASH) |
| Maximum steps of the program |
30 (can be used for multiple nested PCR experiments) |
| Maximum number of loops in the program |
200 (up to 100000 under nested loops)
|
| Time increments/decrease |
1s- 600sCan be used for LongPCR experiments
|
| Temperature increase/decrease |
0.1-10.0℃, can be used for Touchdown experiment
|
| Program pause function |
have |
| Power off data protection |
have |
|
4℃ insulation
|
Infinite insulation time |
| Language setting function |
Chinese/English |
| Real time temperature curve |
Real time recording of the temperature of the hot cover and module, with a single page recording time of 60 minutes, and can be viewed by sliding |
| Language setting function |
Chinese/English |
| GLP report |
Record the running information of each step of the program and export CSV and PDF format reports |
| File encryption function |
Can encrypt individual program files |
| User management permissions |
Third level permissions (administrator, registered user, guest) |
| Email reminder function |
Automatically send email reminders after the experimental program is completed |
| Lock screen function |
Prevent accidental termination of the experimental program |
| Computer online function |
Can simultaneously monitor over 250 instruments online |
| mobile app |
Can simultaneously monitor over 250 instruments online |
| LCD (Liquid Crystal Display) |
10.1 inch,1280×800 for
|
| communication interface |
USB2.0 ,WIFI
|
| Product dimensions |
385mm× 270mm× 255mm (L×W×H)
|
| Net weight |
10kg |
| input power |
100-240VAC , 50/60Hz ,600 W
|