Welcome Customer !

Membership

Help

Jinan Zhongke Electronic Technology Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Products

Integrated design of positive and negative pressure for microbial challenge sealing tester

NegotiableUpdate on 05/07
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

The microbial challenge method sealing tester adopts the principles of positive pressure (pressurization) and negative pressure (vacuuming) testing, and is suitable for microbial challenge method (immersion) packaging system sealing testing of composite bags, blister packaging, oral solid pharmaceutical bottles, glass infusion bottles, glass injection bottles, glass ampoules, oral liquid bottles, hoses and other products used in drug packaging.

Product Details

Integrated design of positive and negative pressure for microbial challenge sealing testerAdopting the principles of positive pressure method (pressurization) and negative pressure method (vacuuming) testing, it is suitable for microbial challenge (immersion) packaging system sealing test of composite bags, blister packaging, oral solid pharmaceutical bottles, glass infusion bottles, glass injection bottles, glass ampoules, oral liquid bottles, hoses and other products used in drug packaging.

Product features:

ØThe integrated design principle of positive and negative pressure can meet the requirements of colorimetric and microbial challenge methods(Immersion) Sealing Test Conditions;

ØIndustrial grade color high-definition touch screen, clear view, menu style interface, sensitive touch, easy to operate;

ØThe experimental process is operated with one click, resulting in high testing efficiency;

ØAutomatic constant pressure air supply ensures that testing can be conducted under preset pressure conditions;

ØThe automatic blowback unloading function makes the end of the test process simple and smooth;

ØPositive pressure, negative pressure, holding time and other data can be preset. Enter the values to enter the test mode;

ØThe instrument has functions such as storing and querying experimental data;

ØLarge scale testing scope to meet users' more testing requirements;

ØUsing internationally renowned pin brand pneumatic components, with excellent sealing performance and more reliable quality;

ØConfigure a micro printer that can directly output and print test results;

ØMulti level user permission management, password login, convenient for user management.


微生物挑战法密封试验仪 正压负压一体设计


Testing principle:

Microbial challenge sealing test, also known as microbial immersion sealing integrity test, is usually conducted simultaneously with simulated filling. By simulating the filling of the culture medium according to the simulation filling verification plan, followed by plugging and capping, the visual inspection is qualified. After sterilization in a validated sterilization cabinet, immerse the sealing surface of the container in a high concentration bacterial solution to ensure that the culture medium inside the sample container fully contacts the inner surface of the seal. The neck of the sample and the outer surface of the seal should be fully immersed in the bacterial suspension. After soaking for a certain period of time, remove it and regularly cultivate it to check for microbial invasion and determine the integrity of the container sealing system. At the same time, a positive control test needs to be conducted to confirm the growth promoting ability of the culture medium. In testing, challenges such as vacuum or overpressure are usually applied,MIT-L20The microbial challenge method sealing tester can meet the requirements of relevant negative pressure vacuum and positive pressure overpressure test conditions.

Test application:

Microbial challenge method (immersion) sealing test for packaging systems of composite bags, blister packaging, oral solid medicine bottles, glass infusion bottles, glass injection bottles, glass ampoules, oral liquid bottles, hoses and other products used in drug packaging.

Technical Indicators:

indicator

parameter

test scope

positive pressure

0-400 KPa (standard) Note: Other ranges can be customized

negative pressure

0~-90 kPa (standard); 0~-97 kPa (optional)

test accuracy

Level 0.5

Pressure resolution

0.01kPa

Compressed air requirements

0.5MPa~0.7Mpa (gas source to be provided by the user)

holding time

0 ~ 99999S

air connection

4 mm polyurethane pipe

Power supply

AC 220V 50Hz

Overall dimensions

Host: 400mm (L) × 300mm (W) × 190mm (H)

Net weight

About 20Kg




Product Configuration:

Standard configuration: host, test tank, micro printer

Optional accessories: non-standard test tank, non-standard test accessories, air compressor, vacuum pump

Note: The air source interface of this machine is a Φ 4mm polyurethane pipe, and the air source should be provided by the user themselves.

Note:Jinan Zhongke Electronic Technology Co., LtdAlways committed to innovation and improvement of product performance and functionality, for this reason, product technical specifications and appearance may also change accordingly, without prior notice. Our company reserves the right of modification and final interpretation.

Integrated design of positive and negative pressure for microbial challenge sealing tester

微生物挑战法密封试验仪 正压负压一体设计