Flyscience2000Infrared oil measuring instrumentCompliant with standards:
●National Environmental Protection Standard "HJ 637-2018 Water Quality - Determination of Petroleum, Animal and Plant Oils - Infrared Spectrometric Method"
●National Environmental Protection Standard "HJ 1077-2019 Determination of Fixed Source Exhaust Gas, Oil Fume and Oil Mist by Infrared Spectrophotometry"
●National Environmental Protection Standard "HJ 1051-2019 Determination of Petroleum in Soil by Infrared Spectroscopy"
Function/Features:
●Instrument qualification: The instrument has a "Certificate of Approval for Measuring Instruments" and a computer software copyright certificate.
●Beautiful and elegant appearance, using ABS engineering plastic casing, resistant to acid and alkali corrosion.
●Adopting high-throughput optical system, integrated optical path design, short optical path length, high energy, small instrument size, light weight, first splitting and then absorbing, meeting the requirements of infrared spectroscopy characteristics, good stability, and high signal-to-noise ratio; Using a precision stepper motor to control the grating, it has wavelength correction function, high wavelength accuracy, and good repeatability;*The design of the colorimetric pool structure is suitable for0.5Any colorimetric dish up to 5 centimeters.
●By using electrically modulated light sources, the heating intensity of the light source is reduced, which is beneficial for system heat dissipation. The lifespan of the light source can reach over 8000 hours. At the same time, due to the absence of mechanical cutting motion components, the instrument structure is simplified and the stability of the instrument is improved.
●Using a mechanical and optical dual calibration system for automatic positioning and wavelength measurement, appropriate concentration oil standard samples are used to automatically calibrate the wavenumbers at 2930cm-1, 2960cm-1, and 3030cm-1, eliminating the influence of human factors and accurately measuring absorbance. ZUI greatly increases instrument stability.
●Specialized data processing system, applicableInfrared spectrophotometry using solvents such as S-316 as extractants in a dual system of tetrachloroethylene and carbon tetrachloride; During the measurement process, the infrared spectrum is continuously scanned, which can clearly distinguish the structure of various oil products and accurately identify interfering substances.
●Equipped with purity detection function for tetrachloroethylene: using a dry 4cm empty quartz colorimetric dish as a reference, in2800 cm-1~3100Between cm-1Using a 4cm quartz colorimetric dish to measure tetrachloroethylene,2930 cm-1The2960 cm-1、 thirtyThe absorbance at 30 cm-1 should not exceed0.34, 0.07, 0. The analysis software automatically determines whether the purity of the tetrachloroethylene reagent is qualified or unqualified.
●Instrument measurement method: The instrument has been newly developed and uses calibration coefficients for direct measurement, which complies with three different national standards: water quality, fixed pollution source exhaust gas, and soil. There is no need to add any proportional coefficients to correct the final value, and no need to create additional standard curves. The final result can be obtained by direct measurement.
●Instrument calibration method: In addition to directly measuring the calibration coefficient, the instrument is also compatible with the function of producing standard curve measurement data. Customers can choose the appropriate calibration and measurement method according to their needs.
●Capable of measuring calibration coefficients: According to HJ638-2018 national standard 8.1, the instrument calibration coefficients X, Y, Z, and F can be calculated using standard solutions of n-hexadecane, isooctane, and benzene.
●Oil analyzer analysis softwareSet measurement parameters according to current national standards, canDirect reading of oil in waterOil fume and oil mist in exhaust gas from fixed pollution sourcesMeasurement results of petroleum in soilNo manual conversion is required.
●dataMeasurement completedAutomatically generatedInspection ReportThe report includes correction coefficient formulas, standard curves and spectra, sample scanning spectra and measurement data, sample measurement parameters, customer information, etc. AdditionallyYou can freely select the display content of the test report, such as sample name, sample volume, extractant volume, dilution ratio, measurement time, sample category, sample concentration, concentration value, absorbance, etc,Various data and corresponding spectra are automatically saved without the need for computer screenshots;When exporting the inspection report, multiple data measurement spectra can be selected and displayed in an overlapping spectrum, or each data can be selected to display its individual measurement spectrum, fully meeting the needs of various customers;
●More adaptability: RS232 and USB communication interfaces are both available; WIN7、 8. 10 computer operating systems are acceptable; You can choose to use wired computers or wireless tablets according to your needs; Equipped with an upgrade interface for the fully automatic oil analyzer, the Flyscience 3000 fully automatic can be directly upgraded without the need for a replacementInfrared oil measuring instrument.
Flyscience2000Infrared oil measuring instrumentTechnical Indicators:
●Baseline stability: Real time automatic adjustment of zero point, automatic calculation of relative absorption peak (eliminating the influence of baseline drift), computer collects both the signal when the light source emits light and the signal when the light source goes out, achieving real-time automatic adjustment of zero point, simplifying operation and improving long-term signal stability. Zero drift is 0.0025A/30min
●Correction coefficient accuracy: Take an appropriate amount of petroleum standard solution, prepare an appropriate concentration of petroleum standard solution with tetrachloroethylene as the solvent, and perform the same steps as the sample measurement. The relative error between the measured value and the standard value is within ± 8%
●Correction coefficient directlyMeasurement value basis: ρ=X·A2930+Y·A2960+Z·(A3030-A2930/F)
●Accuracy of special oil products: measuring 20mg/L n-hexadecane, 20mg/L isooctane, and 100mg/L benzene with an accuracy of ≤± 8%
●Method detection limit (HJ 637-2018): When the volume of the water sample is 500ml and the volume of the extraction solution is 50ml, the detection limit of the water sample is 0.02mg/L
●Instrument detection limit: DL ≤0.02mg/L(Measure 3 times SD of blank solution of tetrachloroethylene 11 times)
●Standard oil accuracy: ≤± 1%
●Repeatability: RSD ≤ 0.2% (11 measurements of 30-60mg/L oil standard sample); RSD ≤ 2% (0.2mg/L water sample)
●Basic measurement range:0.0~1000.0mg/L
●samplemeasurement range:0~100% oil (oil concentration in water sample, adjust extraction ratio)
●Range of absorbance:0.0000-2.0000 AU (i.e. transmittance of 100-1% T))
●Minimum absorbance scale: 0.0001(A)
●Scanning wavenumber range:3400cm-1 to 2400cm-1 (i.e. 294 cm-1)1nm~4167nm)
●Measurement wavelength range: 3200nm~3600nm
●Wave number accuracy and repeatability:±0.5cm-1
●Wavelength accuracy and repeatability: ± 0.5nm
●Wavelength precision:<0.03%(30~60mg/L oil standard sample11 consecutive measurements at a wavelength of 3413nm
●standard curveLinear correlation coefficient:R≥0.9998
●Standard method and measurement time: Infrared spectrophotometry, the instrument has a full wavelength continuous infrared absorption spectrum, full spectrum scanning, 30S/time; Non dispersive infrared photometry, 1 second per measurement.
●external dimensions:465Mm (length) × 315Mm (width) × 165Mm (height)
●host weight:13Kg
●Temperature: -5~45 ℃, relative humidity: 20~80%
●Host power supply power:220 ±10%V,50Hz,50W