- Phone
-
Address
No. 77 Eshan Road, Pudong New Area, Shanghai
Shanghai Jinjia Scientific Instrument Co., Ltd
No. 77 Eshan Road, Pudong New Area, Shanghai
On site installation type conductivity meterHE-200C ESH-01-C-S-ST-Y
Four wire transmission, low concentration
Horiba conductivity meter conductivity meter HE-200C ESH-001Measure the conductivity and temperature in the water sample.
Outdoor on-site installation, different from panel installation of HE480C. According to the USP<645>quality guidelines, it is suitable for monitoring the recovery control of pure water/boiler drainage and online measurement of pharmaceutical water conductivity.
Conductivity: 0-2000 μ s/cm, 0-200.0 ms/cm
(TDS 0 ~ 2000 mg/L)
Temperature: 0~100 ℃
Dealing with harsh conditions during on-site measurement.
Operating temperature range (transmitter): -20~55 ℃
Rated power supply voltage: AC 100~240 V ± 10% 50/60 Hz
Aluminum alloy watch box: protection level IP65
Compliant with CE marking and FCC rules, suitable for use on overseas sites.
host:HE-200C
Split electrode:ESH-001-C-S-ST-Y、ESH-01-C-S-ST-Y
Integrated electrode:ESH-001-L-S-ST-Y-10、
Flow pool:EFA-30、EFA-30-P、
Split cable:CK-10M
Convenient function to display liquid temperature simultaneously (transmission output available).
Respond to various types of temperature compensation.
The self diagnostic function can immediately detect sensor errors.
Environmentally friendly (lead-free) design that can handle WEEE and RoHS.
Rich interfaces (2-channel transmission output, compatible with RS-485, 3-contact output).
Easy to view large display screen.
Easy to operate.
Can store calibration records.
Use thread terminals in the wiring board that will not come off for wiring
parameter
measurement method |
Dual electrode communication method |
||||
|
range (Conductivity measurement and TDS conversion cannot be selected simultaneously) *Range only applicable to hygiene grade sensors (FS-01 series) |
Electrode constant | /cm |
0.01 |
0.1 |
1 |
conductivity |
uS/cm |
0.000~2.000 |
0.000~2.000 |
0.0~200.0 |
|
mS/cm |
0.0000~0.2000 |
0.0000~0.2000 |
0.00~20.00 |
||
TDS conversion |
mg/L |
0.00~2.00 |
0.0~20.0 |
0~200 |
|
temperature |
℃ |
0~100 resolution: 0.01 ℃ | |||
repetitiveness |
Conductivity: ≤± 0.5% F.S. TDS conversion: ≤± 1.5% F.S Temperature: ± 0.1 ℃ (or equivalent input) |
||||
linear |
Conductivity: ≤± 0.5% F.S. TDS conversion: ≤± 1.5% F.S Temperature: ± 0.5 ℃ (or equivalent input) |
||||
Transfer output |
2-4 mA DC input/output insulation type maximum load impedance 900 Ω |
||||
Contact output |
3-point no voltage contact output relay contact SPDT (1c) Contact function: R1, R2: upper limit alarm, lower limit alarm, USP judgment, output signal retention optional FAIL: Fault alarm Output content: conductivity (or TDS conversion value), temperature |
||||
Contact input |
one o'clock Contact form: collector open circuit, no voltage, contact a Contact function: external input for transmission output holding |
||||
communication function |
RS-485 2-wire type, input/output insulation type (transmission output is non insulated) |
||||
temperature compensation |
Range: 0~100 ℃ (extended and calculated for temperatures below 0 ℃ and above 100 ℃) | ||||
ambient temperature |
-20~55℃ |
||||
Temperature compensation component |
Pt1000(0℃) |
||||
| Correction function | Conductivity: Correction coefficient for input electrode constant (parameter input) TDS conversion value: based on specified coefficients (0.30-1.00) conversion Temperature: 1-point calibration compared to the reference thermometer |
||||
Self-diagnostic function |
Sensor diagnostic error, transmitter malfunction |
||||
power supply |
100~240 V AC 50/60 Hz Power consumption: 15 W or lower |
||||
construction |
Outdoor installation type: IP65 protection level |
||||
weight |
About 4.5 kg |
||||
| Electrical certification | CE marking,FCC rules |
||||
Applicable electrodes |
ESH and FS series conductivity sensors (electrode constants of 0.01/cm, 0.1/cm, 1.0/cm) |
||||
Electrode selection ESHSplit style