Features:
• Large capacity ultra purification column, effectively removing ions and organic impurities
Built in 0.2 μ m filter for removing bacteria and particles
Ultra pure water automatically circulates every hour or can be manually cycled with just one click, ensuring high-quality ultra pure water is available at all times
Online monitoring of inlet conductivity and ultra pure water resistivity, with optional temperature compensation/non compensation modes
8-inch color touch control screen, capable of setting and monitoring all modules, and can be operated with water or latex gloves
• Full process monitoring. Real time display of water quality indicators, operating parameters, consumable components (including UV lamps, pumps, valves, etc.) operating status, display of maintenance and alarm information during system operation
A theme interface with illustrations and text, including 9 working languages including Chinese and English
Quantitative water intake (0.1-999 L), with memory function. When the water volume reaches the preset value, the allocation will automatically stop
Purification columns, UV lamps and other consumables accessories are equipped with RFID chips, which are automatically recognized by the host
Automatically store operational data for more than two years, provide traceability basis, and comply with GLP requirements
• Complete and non modifiable water quality reports can be exported via USB, with paperless operation
RephiLink APP can remotely view machine usage related information and manage multiple devices synchronously
• Support online printing, which can instantly print water quality data, etc
• Verifiable standard machine, providing complete IQ/OQ/PQ verification documents, in compliance with GMP/GLP requirements
Specific certification documents can be provided, including calibration certificates, quality certificates, factory inspection certificates, etc
• Three level menu - User, Administrator, Engineer, with different management permissions
• Support laboratory information management system (LIMS) and building management system (BMS), data informatization
• Meet the requirements of FDA 21 CFR Part 11
Magnetic front cover design, tool free installation/replacement of purification column, easy maintenance
The system has a small footprint, easy and flexible maintenance and operation, and simple replacement of consumables
Main applications:
ultrapure water |
|
Performance indicators:
Water inlet requirements | |
water ingress |
Purified water such as EDI water, RO water, distilled water, etc |
inlet water temperature |
5 - 35℃ |
inlet water pressure |
Suggest a pressure greater than 0.2 MPa (2 bar) |
Technical indicators for water production | |
Water production flow rate |
8 L/min (inlet pressure @ 0.2 MPa) |
Electrical resistivity (@ 25 ℃) |
18.2 MΩ.cm |
Total Organic Carbon (TOC)* |
< 10 ppb |
Particles (>0.2 μ m) |
< 1 /ml |
Microorganisms* |
< 1 cfu/ml |
Silicon* |
< 2 ppb |
heavy metal |
< 1 ppb |
Technical Specifications | |
Overall dimensions:Length x Depth x Height (cm) |
57 X 46 X 70 |
host power |
< 600 W |
*The production of water is affected by the inflow conditions, and it is recommended to use pure water that meets the second level water specification in Chinese GB/T 6682-2008 "Specifications and Test Methods for Water Used in Analytical Laboratories" as the inflow
Order Information:
description |
Item Number |
Genie Q Host Kit |
RG0Q01000K |