Hana HI96701 microcomputer residual chlorine analyzer
Product Details


PRODUCT INFORMATION



measuring range0.00 to 5.00 mg/L(ppm)CI2
Adopting innovative optical system design, it has excellent accuracy, repeatability, and measurement time required,HannauniqueCAL Check ™The function allows for performance validation and calibration of instruments using NIST traceable standards, with built-in reaction timers to ensure consistency between sample measurements and user results, and good laboratory practices(GLP)Display the date and time of the last user calibration, battery status indicator light, and automatic shutdown settings
TipsNEWHI97701Residual chlorine analyzer, using menu operation mode, can choose to use powder or water residual chlorineDPDReagents to enhance your multiple choices and reduce usage costs.



TECHNICAL INDEX



technical specifications

HI96701Residual chlorine

measuring range

0.00 to 5.00 mg/L(ppm)CI2

resolution

0.01 mg/L(≤3.50ppm)、0.10 mg/L(>3.50ppm)

accuracy@25°C (77°F)

Reading±3% ± 0.03 mg/L(ppm)

optical system

narrow band interference filtersilicon photocellTungsten lamp @ 525 nm

Method standards

USEPAmethod330.5, DPDmethod

Power Mode

1 x 9V battery

Automatic shutdown

In measurement mode, measurement mode10In minutes, or in calibration mode, it will automatically shut down after one hour of no use

application environment

0 to 50 °C(32 to 122°F ); RH-max95%(No condensation)

Dimension

Host size:192 x 104 x 69 mm (7.6 x 4.1 x 2.7”)、Host weight:320 g (11.3 oz.)

Exclusive attachments

HI731333Glass cuvette(HI731331+HI731335)Specification: Outer diameter24.6Mm (inner diameter)22mm),4Set/Group

HI93701-01 Customized residual chlorine (free chlorine) DPD reagent prediction times100order

HI93701-03 Customized residual chlorine (free chlorine) DPD reagent prediction times300order

HI96701-11 Residual chlorineNISTMeasurement curve calibration group, standard value@25ºC:0.00 and1.00 ± 0.03mg / L(ppm)CI2




Standard configuration


HI96701

Host, HI93701-01 dedicated residual chlorine DPD reagent, HI731333 customized glass colorimetric dish (2 pieces), HI731318 customized glass colorimetric dish cleaning cloth (2 pieces), Chinese and English user manual, HI721006 customized instrument carrying case [Warm reminder: To verify instrument performance, please order HI96701-11 residual chlorine NIST measurement curve calibration group]

HI96701C

Host, HI93701-01 dedicated residual chlorine DPD reagent, HI96701-11 residual chlorine NIST measurement curve calibration group, HI731333 customized dedicated glass colorimetric dish (2 pieces), HI731318 dedicated glass colorimetric dish cleaning cloth (2 pieces), Chinese and English user manual, HI721018 customized dedicated instrument carrying case