Hamilton Dissolved Oxygen Electrode 10118255-1411

NegotiableUpdate on 01/29
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
Hamilton Dissolved Oxygen Electrode 10118255-1411 is an optical dissolved oxygen (DO) sensor that integrates optoelectronic components and has complete measurement and self diagnostic capabilities. It adopts RS485 Modbus and ECS (electrochemical signal simulation) interfaces, making it an ideal choice for fermentation processes in the biopharmaceutical, biotechnology, and food and beverage industries.
Product Details

Hamilton Dissolved Oxygen Electrode 10118255-1411

Product Overview

  • Product Name: VisiTherm RS485-ECS 325 H0

  • Order Number: 10118255-1411

  • Product Description: This is an optical dissolved oxygen (DO) sensor that integrates optoelectronic components and has complete measurement and self diagnostic capabilities. It adopts RS485 Modbus and ECS (electrochemical signal simulation) interfaces, making it an ideal choice for fermentation processes in the biopharmaceutical, biotechnology, and food and beverage industries.

  • Core advantages:

    • Optical measurement: No need to replace electrolyte or polarographic film, extremely low maintenance.

    • Dual interface: Supports both modern digital communication (RS485) and analog signals from traditional electrochemical electrodes (ECS), making it convenient for upgrading old devices.

    • No flow rate dependence: can accurately measure even in stationary liquids.


Hamilton Dissolved Oxygen Electrode 10118255-1411Technical Specifications Table

Parameter item Value/Description
Sensor length (a-length) 325 mm
Type of Sensor Cap ODO Cap H0 (stainless steel 1.4435+silicone, sanitary grade design)
Measuring principle Optical fluorescence quenching technology (no electrolyte required, no oxygen consumption)
Measuring Range

4 ppb to 25 ppm (liquid concentration)


0.1 to 625 mbar (oxygen partial pressure pO2)


0.05 to 300% sat (saturation)

Accuracy

At 25 ° C:


1 ± 0.05 %-vol


21 ± 0.2 %-vol

Drift <1% per week at room temperature
Temperature Sensor Built in NTC 22 kOhm
Analog Interface ECS (Simulating the nanoampere level current signal of traditional polarographic electrodes)
Digital Interface RS485 Modbus (supports up to 32 addresses)
Baud Rate 4800, 9600, 19200, 38400, 57600, 115200 Bd
Operating Voltage 10 to 27 VDC (maximum power of 1.5 W)
Operating temperature range (Op. Temp)

-20 to 140 ° C


(Note: Effective DO readings are only provided below 85 ° C, and high temperatures are mainly used for sterilization)

Pressure Range -1 to 12 bar (gauge pressure)
Protection Rating IP68 waterproof
Sterilization and cleaning

Autoclave sterilization


SIP online steam sterilization (max 140 ° C)


CIP online cleaning

Wetted Parts Material

Stainless steel 1.4435


Silicone (compliant with FDA 21 CFR 177.2600, USP Class VI standards)

surface finish Ra < 0.4 µm (N5)
response time T98%<30 seconds (at 25 ° C)
Explosion-proof certification None (This model is not ATEX/IECEx explosion-proof version)