The HG-C hydride generator can not only be used in conjunction with atomic absorption, but also with ICP plasma. And it can be used in combination with liquid phase and ultraviolet for morphological analysis. It has surpassed technology.

Product Details

HG-C hydride generatorNot only can it be used in conjunction with atomic absorption, but it can also be used in conjunction with ICP plasma. And it can be used in combination with liquid phase and ultraviolet for morphological analysis. It has surpassed technology. At present, our company's hydride generator is used for atomic absorption in China. And foreign atomic absorption instrument manufacturers and ICP hydride generator plasma manufacturers are gradually using our company's hydride to replace imported hydride generators.

1. Adopting peristaltic pump technology, the inlet flow rate can be flexibly set.

2. Prepare a new gas-liquid separator to automatically remove waste liquid.

3. Optimize gas path control, limit baseline noise suppression, and improve signal-to-noise ratio.

4. * Atomizer with automatic temperature setting and excellent temperature control accuracy.

5. Controllable continuous injection, using any atomic absorption instrument for detection.

HG-C hydride generatorWorking Principle:

Some elements in aqueous solution, such as As, Se, Sn, or Bi, are mixed with HCl and NaBH4 solvents through a peristaltic pump and sent into a reaction loop to generate hydrides.
The generated mixture consists of hydrides, hydrogen gas, water vapor, and residual reagents, which are sent to a gas-liquid separator through a carrier gas. In the gas-liquid separator, the separated gas phase is sent into a quartz tube through a carrier gas. Heating a T-shaped quartz tube with acetylene combustion to thermally decompose hydrides. Laboratory equipment conditions in aqueous solutions in the presence of reducing agents.
The laboratory should be equipped with exhaust equipment, and the size of the exhaust volume should be able to gently suck up a slightly larger newspaper attached to the exhaust outlet. Excessive air volume can affect the stability of flames; The air volume is too small and unstable, and harmful gases cannot be discharged.
The test bench is sturdy and stable, with a flat surface. The space left around the test bench should not be less than 0.5 meters for operation and maintenance.
Gas source preparation conditions: When using flame method, acetylene steel cylinders are required, with a purity greater than 98% and an outlet pressure of 0.08MPa. Acetylene gas cylinders should not be placed in the main engine room, but should be placed in a safe and smoke-free room closest to the main engine. However, the main engine room must be equipped with a gas circuit switch valve to quickly cut off the gas in case of an accident.