HCGC-280 Gas Chromatography Analysis System for Transformer Oil

NegotiableUpdate on 06/07
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

HCGC-280 Gas Chromatography Analysis System for Transformer Oil

Product Details

Detailed introduction:

HCGC-280It is an economical and applicable gas chromatography analysis system for dissolved gases in transformer oil launched by Shenzhen Huachang Haitai Technology Co., Ltd. The system uses a domestically produced host (three detectors) and a dedicated analysis process to completely separate the insulating oil in one injection7Dissolved gas components (includingH2COCO2CH4C2H4C2H6C2H2), can meetThe need for chromatographic analysis of dissolved gases in transformer oil in the power system.

Technical features:

l Large screen LCD display

l Chinese Menu

l key operation

l digitizeFIDElectronic zeroing technology enhances the anti-interference ability of instruments

Using electronic zeroing technology to achieveFIDElectronic regulation of output signal level. Due to the fact that electronic adjustment can be directly installed on the signal channel of the circuit board, the electronic noise suppression capability is improved, and there is no interference signal when connecting mechanical potentiometers. Due to the reduction of parasitic parameters, it has strong anti-interference ability.

l Thermal conductivity tungsten wire gas cutoff protection function

HCGC-280Equipped with intelligent thermal conductivity cut-off protection function, when the carrier gas pressure is below the safety threshold, the system will automatically cut off the thermal conductivity current to protect the thermal conductivity detector from damage.

Technical parameters:

1 Small detection concentration (unit)µL/L):

H2

CO

CO2

CH4

C2H4

C2H6

C2H2

2

2

2

0.1

0.1

0.1

0.1

2 Qualitative repeatability: deviation≤1

3 Quantitative repeatability: deviation≤3

4 Execution standards:

5 GB/T 17623-1998Gas chromatography method for determining the content of dissolved gas components in insulating oil
GB/T 7252-2001
Guidelines for Analysis and Determination of Dissolved Gases in Transformer Oil

6 Technical specifications of detector:

Thermal conductivity detector(TCD

Hydrogen flame ionization detector(FID

Adopting a semi diffusive structure

Cylindrical collection electrode structure design, quartz nozzle, extremely responsive

The power supply adopts constant current control mode

Automatic ignition; Stable time: Ten minutes

Sensitivity: ≥2500mV·mlmg(Hexadecane)/Isooctane)

Detection limit: ≤8×10-12gs(Hexadecane)/Isooctane)

Baseline noise: ≤20μV

Baseline noise: ≤2×10-13A

Baseline drift: ≤100μV/30min

Baseline drift: ≤2×10-12A/30min

Line Sex: ≥104

Linear: ≥106

7 Temperature control range: above room temperature8~399 Accuracy ±0.1