Overview

The high-performance absolute pressure transmitter and gauge pressure transmitter EJX610A and EJX630A use monocrystalline silicon resonant sensors, suitable for pressure measurement of liquids, gases, and vapors. Multi sensor technology provides

Product Details

The high-performance absolute pressure transmitter and gauge pressure transmitter EJX610A and EJX630A use monocrystalline silicon resonant sensors, suitable for pressure measurement of liquids, gases, and vapors.
Multi sensor technology provides advanced diagnostic functions that can detect abnormalities such as pulse line blockage or damage to heat tracing pipelines.
The pressure transmitter EJX630A can be selected with a high-precision option (± 0.025%).
The EJX series pressure transmitter supports BRAN, HART, and FOUNDATION ™ Fieldbus communication protocol.

specifications

EJX610A and EJX630A
(The following measurement results are all data from EJX610A)
A membrane box B membrane box C membrane box D membrane box
measuring range EJX610A 0~200kPa
(0~29psi)
0~2MPa
(0~290psi)
0~10MPa
(0~1450psi)
0~70MPa
(0~10150psi)
EJX630A -100~200kPa
(-14.5~29psi)
-0.1~2MPa
(-14.5~290psi)
-0.1~10MPa
(-14.5~1450psi)
-0.1~70MPa
(-14.5~10150psi)
Measurement range 2~200kPa
(0.3~29psi)
0.01~2MPa
(1.5~290psi)
0.05~10MPa
(7.3~1450psi)
0.35~70MPa
(50.8~10150psi)
accuracy
(EJX630A can choose high-precision option)
±0.04% ±0.04% ±0.04% ±0.04%
protection grade IP67,NEMA 4X
Explosion proof specifications FM,CENELEC ATEX,CSA,IECEx
output signal 4-20mA DC or FOUNDATION ™ Fieldbus, two-wire digital communication
power supply BRAN and HART: 10.5~42V DC
(Intrinsic safety type: 10.5-30V DC)
Fieldbus: 9-32V DC
(Intrinsic Safety Entity Mode: 9-24V DC, Intrinsic Safety FISCO Mode: 9-17.5V DC)
ambient temperature -40-85 ℃ (-40-185 ℉) (ordinary type)
-30-80 ℃ (-22-176 ℉) (with built-in indicator)
process temperature -40-120 ℃ (-40-248 ℉) (ordinary type)
Maximum overvoltage 4MPa
(580psi)
16MPa
(2300psi)
25MPa
(3600psi)
105MPa
(15200psi)
install Direct installation/2-inch pipe installation
Material of liquid receiving components Capsule Hastelloy C-276
Process connector 316L SST, Hastelloy C-276
casing Aluminum alloy or ASTM CF-8M stainless steel (optional)