Gas concentration detector

NegotiableUpdate on 05/07
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

The LGDS-1500 gas concentration detector is a key project achievement of national environmental monitoring. It uses laser sensing principle to achieve online monitoring of methane gas and has the characteristics of fast, accurate and stable. This product can be widely used for gas safety monitoring in landfill gas power plants, petrochemicals, coal mines, and other scenarios, and can achieve remote data transmission.

Product Details

LGDS-1500Gas concentration detectorProduct advantages:

Core technology control, using laser sensing core technology, with outstanding advantages of "fast", "accurate", and "stable";

Multi level pretreatment measures effectively filter impurities and water vapor, and include automatic drainage function, resulting in a longer product lifespan;

PLC control output, capable of integrating multiple functions;

Multi channel integration enables multi-point integration testing solutions, facilitating on-site multi-point parameter testing and saving costs;

Large capacity data storage, with functions such as historical data storage, query, and export;

High cost performance solution, advantages of mass production and low cost;

Multi parameter measurement, simultaneous measurement of gas composition and humidity scheme, as well as supporting technology upgrades to display flow rate, etc;

Large capacity historical data storage, capable of storing data for up to 15 years.

LGDS-1500Gas concentration detectorProduct Features:

Functional characteristics of the whole machine

Real time digital display online monitoring of multiple gases and humidity

Automatic drainage, no manual operation required

High reliability, intrinsic anti-interference (non methane gas has no reaction)

Automatic pumping and measurement

Intrinsically zero drift free, no calibration required during service life (CH4)

Comes with oil and gas filtration, water vapor filtration, and impurity filtration

Explosion proof design

Molecular sieve plug-in operation, easy to replace

Waterproof and dustproof design

Multi channel compatible design, greatly saving customer costs

Typical application scenarios of laser gas online monitoring device:

Before desulfurization (gas types can be NH3, H2, etc.)

component content

measurement range

Test Principle

Precision

methane

CH4:0-100% VOL

Laser principle

±1%

carbon dioxide

CO2: 0-100% VOL

Infrared principle

±2%

oxygen

O2 : 0-30% VOL

electrochemistry

±1%

hydrogen sulfide

H2S: 0-10000 ppm

electrochemistry

±2%

After desulfurization (gas types can be NH3, H2, etc.)

component content

measurement range

Test Principle

Precision

methane

CH4:0-100% VOL

Laser principle

±1%

carbon dioxide

CO2: 0-100% VOL

Infrared principle

±2%

oxygen

O2 : 0-30% VOL

electrochemistry

±1%

hydrogen sulfide

H2S: 0-100 ppm

electrochemistry

±1%

Humidity measurement

measurement range

0~100% RH biogas specific corrosion-resistant imported sensor

accuracy

At 0...+40 ° C:

±1.5 %RH (0 … 90 %RH)

±2.5 %RH (90 … 100 %RH)

At -40... 0 ° C,+40...+80 ° C

±3.0 %RH (0 … 90 %RH)

±4.0 %RH (90 … 100 %RH)

stability

+-2% RH (over 2 years)

Technical parameters:

Configuration parameters of LGDS-1500 laser gas online monitoring device

Technical parameters (capable of online monitoring of gases such as CH4, CO2, H2S, O2, etc.)

gas type

CH4

CO2

H2S

O2

Measuring principle

Laser type

Infrared (NDIR)

electrochemistry

electrochemistry

Sampling method

extractive

extractive

extractive

extractive

measurement range

0-100%

0-100%

0-100,0-1200,

0-5000, 0-10000ppm

0-30%

measurement accuracy

0-1%: ±0.04%

>1.00%: ±1%

±2%F.S.

±1%F.S.; ±2%F.S.

±1%F.S.

response time

T90<10S

T90<30S

T90<30S

T90<30S

Complete machine configuration parameters

working power supply

100-240VAC

Explosion-proof rating

Exd II BT4 Gb

output method

4-20mA/RS485

Operating Temperature

-20℃~60℃

Waterproof and dustproof

IP65

Output channel

Single channel/dual channel

Storage temperature

-40℃~85℃

Overall dimensions

650*600*250

connection method

shielded cable

Ambient Humidity

0-95%

Installation method

Wall mounted/vertical

pressure control

Optional two-stage pressure reduction