- Phone
-
Address
Workshop 4, 3rd Floor, Building 6, No. 775 Chihua Road, Fengxian District, Shanghai
Shanghai Jiehan Experimental Equipment Co., Ltd
Workshop 4, 3rd Floor, Building 6, No. 775 Chihua Road, Fengxian District, Shanghai
1)Fully automatic anaerobic incubatorbox
Box:Material304 Stainless steel, thickness 3mm
Inner surface: stainless steel brushed treatment Outer surface: painted, white
Equipment main box Length:1220mmDepth:760mm; Height:900mm
Front window:Tilted window, transparent tempered safety glass, thickness8 mm
glove opening:The material is anodized aluminum, processed from solid aluminum, and made of ethylene propylene diene monomerOThe ring is highly corrosion-resistant,diameter220mm
gloveImported butyl rubber, brand: USAnorthsafetythickness0.4mm, diameter8”, Length32”
filter:specification0.3Micrometer,1A gas inlet,1A gas outlet.
Shelves:Stainless steel material,2One, hanging on the back panel, adjustable up and down
Box lighting:No UV white light, installed above each glass window
interfacebackup interface3One,DN 40 KFpower interface 1 A(220V)
2)bigrotundityTransition cabin
Size: Diameter385mm, Length 600mm
Material304Stainless steel surface: The inner surface is treated with wire drawing, and the outer surface is painted
Attachment
Sliding tray,304Stainless steel
Cabin door: double door, anodized aluminum material, thickness10mmVertical operation with lifting mechanism
Pressure gauge: simulated display
control
Electromagnetic valve touch screen automatic operation
3)Small Ferry Cabin
Transition cabin
Size: Diameter150mm, Length300mmEnter the glove box section for length100 mm.
Material304 Stainless steel surface: The inner surface is treated with wire drawing, and the outer surface is painted
Attachment
Cabin door: double door, threaded locking pressure gauge: simulated display
control
Manual valve manual operation
4)Control system:
function:Including self diagnosis, power-off self starting characteristics, with pressure control and adaptive functions; Automatic control, loop control, password protection; Unit control adoptsPLCtouch screen.
Pressure control:Control the working pressure of the control box and transition chamber+/- 10mbarCan be freely set internally, beyond+/- 12mbarSystem automatic protection.
Foot pedal:Control the pressure of the box for easy operation of boosting and lowering pressure
5)Display system:adoptPLCtouch screenSmart700Display the operating status, box pressure, system records, etc.
6)Vacuum system control situation
The vacuum pump can be operated manually or throughPLCStart, traffic8m3/hIt can evacuate the transition chamber and maintain pressure balance in the box, with the maximum vacuum degree of the vacuum pump≤2x10-1pa
7)Fully automatic anaerobic incubatorAttachment
Oxygen probe
Zirconia probes imported from the UK, throughPLCdisplay
Measuring range:0~1000ppm High precision, no drift, accuracy:+/-1%ppm
Display: The output of the analyzer is connected toPLC,The detection value is displayed on the touch screen, and the alarm value can be set.
Control:The gas circulation system can be automatically controlled according to user settings to achieve automatic circulation.