Welcome Customer !

Membership

Help

Kanglaituo Testing Equipment (Suzhou) Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Products

Kanglaituo Testing Equipment (Suzhou) Co., Ltd

  • E-mail

    xsf@kanglaituo.com

  • Phone

    15962643842

  • Address

    151 Chenfeng East Road, Yushan Town, Kunshan City, Suzhou City

Contact Now

Formaldehyde VOC testing chamber

NegotiableUpdate on 02/08
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
The formaldehyde VOC testing chamber is suitable for industries such as scientific research, quality inspection, furniture, home furnishings, electricians, electronics, automobiles, materials, chemicals, and home appliances.
Product Details

Product Name:Formaldehyde VOC testing chamber(Formaldehyde emission detection release box)

Formaldehyde VOC testing chamberApplicable industries: scientific research, quality inspection, furniture, home furnishings, electricians, electronics, automobiles, materials, chemicals, household appliances.

Product use: Suitable for measuring formaldehyde emissions from various indoor decorative materials such as artificial boards and composite wooden floors, constant temperature and humidity balance treatment of wood or artificial boards, and detection of volatile harmful gases in other building materials. Scientific and reasonable design and manufacturing, easy to use, are products of production enterprises, research institutions, technical supervision departments, colleges and universities, and other units.
operating procedure
1. Preparation work
Environmental Requirements
Ensure that the test chamber is placed in a well ventilated environment without strong electromagnetic interference.
It is recommended to control the room temperature at 20-30 ℃ and the humidity at ≤ 70% RH.
Equipment inspection
Check whether the power supply, gas circuit (such as connecting external gas), and sealing of the test box are normal.
Confirm that the internal sensors (temperature and humidity, VOC sensors) have been calibrated and are within their validity period.
Sample preparation
Cut or install samples according to testing standards (such as GB/T 18883, ISO 16000, etc.) to avoid contamination.
If testing materials (such as furniture, boards), pre-treatment (such as edge sealing, aging) should be carried out according to standard requirements.
2. Device startup and parameter settings
power on
Turn on the power, start the control panel, and preheat the equipment to a stable state (usually takes more than 30 minutes).
Parameter Settings
Set the following parameters according to the testing standards:
Temperature: Commonly set at 23 ℃ or 28 ℃ (as required by GB/T 18883 for 23 ℃).
Humidity: usually set at 50% RH.
Air exchange rate: such as 1.0 times/hour (adjusted according to standards).
Testing time: generally ranging from 24 hours to 28 days, depending on the testing purpose.
Background value calibration
Run the test chamber without load to detect the background formaldehyde/VOC concentration and ensure that the background value is below the testing limit.
3. Sample placement
Place the sample evenly inside the test chamber to avoid obstructing the airflow channel.
Maintain a minimum distance of 10cm between the sample and the box wall to ensure even air circulation.
Record the sample number, location, and placement time.
4. Start testing
Close the box door, start the testing program, and the equipment will automatically control temperature, humidity, and air exchange.
Real time monitoring of data (such as temperature, humidity, VOC concentration), and immediate suspension of inspection upon discovery of abnormalities.
5. Data collection and analysis
Regularly collect gas samples using built-in sensors or external sampling devices (such as Tenax tubes, DNPH sampling tubes).
Analyze the concentration of formaldehyde/VOC using a gas chromatograph (GC-MS/HPLC) and calculate the release rate.
Export data and generate a report, comparing standard limits (such as indoor air quality standards).
6. Test completed
Turn off the device power, open the box door for ventilation until the residual gas dissipates.
Clean the inner walls of the box and sample rack to avoid cross contamination.
Maintenance and Care
1. Daily maintenance
clean
After each test, clean the residue inside the box and use a dust-free cloth dipped in neutral cleaning agent to wipe the inner wall.
Regularly clean or replace the air filter (it is recommended to check it once a month).
Sensor maintenance
Formaldehyde/VOC sensors should be calibrated every 3-6 months, or according to the manufacturer's recommended cycle.
Avoid the sensor from coming into contact with high concentration gases or corrosive substances.
leak test
Regularly check whether the door seal is aging and replace it if necessary to ensure airtightness.
2. Regular maintenance
System calibration
Every year, professional technicians perform overall calibration on the temperature and humidity control system and gas flow meter.
consumable replacement
Replace consumables such as activated carbon filters and desiccants (the cycle depends on the frequency of use).
Software Update
Upgrade device control software to ensure compatibility with the latest testing standards.
3. Long term discontinuation
Disconnect the power supply, evacuate the internal gas, and keep the box door slightly open to prevent mold growth.
Seal key components (such as sensors) against moisture.
Precautions
safety protection
Wear protective gloves and masks during operation to avoid direct contact with harmful gases.
When testing high concentration samples, it is necessary to connect an external exhaust system.
environmental interference
Avoid using volatile substances such as solvents and perfume near the test chamber.
Fault handling
common problem:
Temperature and humidity out of control: Check if the sensor, humidifier/dehumidification module are functioning properly.
Data abnormality: recalibrate the sensor or check for blockage in the sampling pipeline.
The device cannot start: Check for power, fuse, or control system faults.
Complex faults require contacting the manufacturer's technical support.
frequently asked questions
Q1: Poor repeatability of test results?
Possible reasons: uneven sample, temperature and humidity fluctuations, sensor drift.
Solution: Standardize the sample preparation process and recalibrate the sensor.
Q2: Is the background value too high?
Possible reasons: box contamination or filter failure.
Solution: Clean the box and replace the filter, run it without load until the background value meets the standard.
Q3: How to extend the lifespan of equipment?
Regular maintenance, avoiding overload operation, and maintaining a clean and dry environment.
By standardizing operations and regular maintenance, the accuracy of formaldehyde/VOC testing chamber data can be ensured and the service life of the equipment can be extended.

sex
can
point
mark
temperature range 15~60℃
temperature fluctuation ≤±0.5℃
temperature uniformity ≤2℃
Temperature measurement accuracy ≤0.1℃
temperature deviation ≤±2.0℃
Temperature stabilization time ≤ 40 minutes
humidity range 25~80%RH
Humidity fluctuation ≤±2%RH
Humidity measurement accuracy 0.5%RH
Humidity stabilization time ≤ 80 minutes
Humidity deviation ≤±3.0%RH
Studio detachment temperature 60℃
airtightness Measurement accuracy level 0.5, VL<5% x gas supply rate Vs, equipment recovery rate:>80%, relative positive pressure maintenance: 10 ± 5 Pa, pressure monitoring inside the box, analog digital display micro pressure sensor, touch screen online pressure display inside the box
Air exchange rate 0.5-3 times/hour (adjustable)
Flow meter accuracy Gas mass flowmeter, ± 1% of full range (including linearity); 0~5L/min
traffic display Micro electronic gas mass flowmeter with touch screen online flow display (adjustable flow range)
Studio wind speed The wind speed at the geometric center point inside the box is 0.1-0.3m/s. (Wind speed adjustable)
Background formaldehyde concentration <0.006mg/m³
noise ≤ 55db (A-level sound level)
make
use
material
material
Exterior materials High quality cold-rolled steel plate with spray coating treatment, plate thickness 1.5mm (color 7035 gray)
Interior materials Stainless steel mirror panel (SUS 304 2B CP grinding processing), plate thickness 1.5mm, fully sealed welded, passivated and polished treatment
insulation layer T70mm (insulation material is formaldehyde free ultra-fine glass wool)
Door observation window W380mm * H480mm
inspection hole Place a diameter of 50mm in the right box
sampling port 2 pieces with a diameter of 6mm, located on the right side of the box
Sampler tray Equipped with a 9-hole power socket
Sample holder SUS 304, Stainless steel mirror sheet metal production
nucleus
heart
match
place
controller Touch screen MCGS, industrial grade, true color 10 "TFT LCD display widescreen
compressor Fully enclosed single-stage compression refrigeration, air-cooled, Danfoss, Denmark
humidifier Volatile electric humidifier, stainless steel water tank, with water shortage reminder function
pressure sensor QBM 2030 Siemens Precision Level 0.5
Safety protection device Prevent overcurrent and short circuit, overvoltage/undervoltage, protect the air supply fan, protect the fan from overheating, protect the compressor from overpressure, overheating and overcurrent, protect the controller from internal overheating, and add an independent temperature rise protector. When the water supply tank is short of water, it will automatically remind and install an air burning prevention device to prevent humidifier from air burning
Water supply conditions Use purified water or distilled water with a conductivity of ≤ 10 μ S/cm2
power supply conditions AC220V 1 Φ 3W 50Hz power: 2.5KW
Usage environment Temperature:+5~+35 ℃; Humidity: not greater than 85% RH; Atmospheric pressure: 80KPa~106KPa