E-mail
sales@open17.com
-
Phone
13817271790
-
Address
Room 205, Building 2, Xindong Real Estate Building, 388 Xinfu Road, Minhang District, Shanghai
Shanghai Yongshi Electronics Co., Ltd
sales@open17.com
13817271790
Room 205, Building 2, Xindong Real Estate Building, 388 Xinfu Road, Minhang District, Shanghai
Product Features
Three in one function: simultaneous measurement of smoke concentration, pressure, and smoke temperature
Compact space can accommodate up to 6 gas sensors simultaneously
• Standard configuration: O2, optional: CO/H2, NO, COhigh, NO2, SO2, one to five types
Equipped with dual air pumps, the backup fresh air pump can automatically turn on to flush the CO/H2 sensor, avoiding sensor poisoning
The large screen displays measurement values in 10 rows, and graphical display can also be selected
Users can customize the quick measurement mode to simplify the measurement steps
• Calculation parameters: thermal parameters such as CO2, NOx, combustion efficiency Effi, flue gas loss value qA, excess air value λ, etc
product description
Technical Parameter
| Smoke temperature/temperature difference | Measurement range: 0 ℃~+1000 ℃ Resolution: 1 ℃ Accuracy: ± 2 ° C+1 bit (-20 ° C~0 ° C) ± 1 °C (0°C ~ 200 °C) ± 0.5 % (﹥200 °C) |
| ambient temperature | Measurement range: -20 ℃~+200 ℃ Resolution: 0.1 ℃ Accuracy: ± 2 ° C+1 bit (-20 ° C~0 ° C) ± 0.5 ° C+1 position (0 ° C~200 ° C) |
| Pressure/Differential Pressure |
Measurement range: standard ± 70 hPa/maximum ± 130 hPa |
| O2 | Measurement range: 0~21 Vol. -% Resolution: 0.1 Vol. -% Accuracy: ± 0.2 Vol. -% |
| CO2count | Measurement range: 0~CO2 max Resolution: 0.1 Vol. -% Accuracy: ± 0.2 Vol. -% |
| CO/H2 | Measurement range: 0~4000 ppm Resolution: 1 ppm Accuracy: ± 5 ppm (≤ 50 ppm) ± 5 % ( >50 ppm) |
| NO | Measurement range: 0~2000ppm (standard)/5000 ppm (maximum) Accuracy: ± 5 ppm (≤ 100 ppm) ±5 % (﹥100 ppm) |
| NO2 | Measurement range: 0~200ppm (standard)/1000 ppm (maximum) Resolution: 1 ppm Accuracy: ± 10 ppm (≤ 50 ppm) ±10 %(>50 ppm) |
| SO2 | Measurement range: 0~2000ppm (standard)/5000 ppm (maximum) Resolution: 1 ppm Accuracy: ± 10 ppm (≤ 200 ppm) ±5 % (﹥200 ppm) |
| COhigh | Measurement range: 0~2 Vol. -% (standard)/10 Vol. -% (maximum) Resolution: 0.01 Vol. -% Accuracy: ± 5% ± 1 digit |
| Weight (including shell) | About 750g |
| Dimensions (W x H x D) | 95 x 215 x 45 mm |
| display | lcd screen |
| pressure port | Ø 7 mm |
| Gas interface | Ø 8 mm |
| power supply | Nickel hydrogen battery (6 V/2 Ah) |
| Operation time (power-saving mode) | 10 h |
| storage | 100 sets of data |
| communication interface | USB, Infrared, Bluetooth |
| Certification standards |
BlmSchV, KÜO (TÜV ByRgG 246), DIN EN50379-2 |