E-mail
sales@open17.com
-
Phone
13817271790
-
Address
Room 205, Building 2, Xindong Real Estate Building, 388 Xinfu Road, Minhang District, Shanghai
Shanghai Yongshi Electronics Co., Ltd
sales@open17.com
13817271790
Room 205, Building 2, Xindong Real Estate Building, 388 Xinfu Road, Minhang District, Shanghai
Product Features
• High end configuration: Integrated Peltier electronic cooler and infrared printer, built-in peristaltic pump automatically eliminates condensed water
• Complete sampling: Optional 3m or 5m electric heating sampling tube, especially suitable for humid flue gas conditions
CO high range extension: The measurement range of the COhigh sensor can be automatically extended to 100000ppm
• Standard configuration: O2、 COH2、 NO, Optional: COhigh, NO2、 SO2One to three types
Equipped with dual air pumps, the backup fresh air pump can automatically start flushing CO/H2Sensors, to avoid sensor poisoning
• Calculation parameter: CO2、 NOx、 Thermal parameters such as combustion efficiency Effi, flue gas loss value qA, and excess air value λ
product description
M650 is a portable flue gas analyzer, with the core analysis components built into a sturdy enclosure with a protection level of IP65. It also integrates a thermal printer and a Peltier electronic cooler. The tested gas is cooled by the Peltier cooler and filtered out of dust by a fine filter before entering the electrochemical sensor inside the device for concentration analysis. The LCD backlit display screen can simultaneously display 10 rows of measurement values. The instrument has measurement value graphic display and flue center temperature search function. The built-in second air flushing pump can automatically turn on to protect the CO/H2 sensor and extend its service life. Automatically display maintenance information for easy maintenance.
The M650 flue gas analysis system integrates a Peltier cooler, so it is mainly used in long-term measurement scenarios (automatic removal of condensed water) and for measuring high concentrations of NO2 and SO2. Because NO2 and SO2 are easily soluble in water, we must eliminate the influence of low measurement results caused by the solubility of the tested gas in water during measurement. So when detecting NO2 and SO2, it is necessary to use a heat tracing sampling tube and a gas cooler. By combining these devices, we can avoid the occurrence of the above situation. The heated gas is cooled by a cooler to remove condensed water, achieving gas water separation.
Technical Parameter
| Smoke temperature/temperature difference | Measurement range: 0 ℃~+1000 ℃ Resolution: 1 ℃ Accuracy: ± 2 ° C+1 bit (-20 ° C~0 ° C) ± 1 °C (0°C ~ 200 °C) ± 0.5 % (﹥200 °C) |
| ambient temperature | Measurement range: -20 ℃~+200 ℃ Resolution: 0.1 ℃ Accuracy: ± 2 ° C+1 bit (-20 ° C~0 ° C) ± 0.5 ° C+1 position (0 ° C~200 ° C) |
| Pressure/Differential Pressure | Measurement range: standard ± 70 hPa/maximum ± 150 hPa Resolution: 0.01 hPa (0 hPa~± 20.99 hPa) 0.1 hpa(± 21.0 hPa ~ ±130.0 hPa) Accuracy: ± 0.02 hPa+1 digit (0 hPa~± 2.00 hPa) ± 1 % (± 2.01 hPa ~ ± 70.0 hPa) ± 2 % (± 70.1 hPa ~ ± 130.0 hPa) |
| O2 | Measurement range: 0~21 Vol. -% Resolution: 0.1 Vol. -% Accuracy: ± 0.2 Vol. -% |
| CO2count | Measurement range: 0~CO2 max Resolution: 0.1 Vol. -% Accuracy: ± 0.2 Vol. -% |
| CO/H2 | Measurement range: 0~4000 ppm Resolution: 1 ppm Accuracy: ± 5 ppm (≤ 50 ppm) ± 5 % (﹥50 ppm) |
| NO | Measurement range: 0~2000ppm (standard)/5000 ppm (maximum) Resolution: 1 ppm Accuracy: ± 5 ppm (≤ 100 ppm) ±5 % (﹥100 ppm) |
| NO2 | Measurement range: 0~200ppm (standard)/1000 ppm (maximum) Resolution: 1 ppm Accuracy: ± 10 ppm (≤ 50 ppm) ±10 %(>50 ppm) |
| SO2 | Measurement range: 0~2000ppm (standard)/5000 ppm (maximum) Resolution: 1 ppm Accuracy: ± 10 ppm (≤ 200 ppm) ±5 % (﹥200 ppm) |
| COhigh | Measurement range: 0~4 Vol. -% (standard)/4 Vol. -% (maximum) Resolution: 0.01 Vol. -% Accuracy: ± 5% ± 1 digit |
| Weight (including shell) | About 6800g |
| Dimensions (W x H x D) | 410 x 175 x 330 mm |
| display | LCD screen, 60 x 80 mm |
| protection grade | IP65 (when closed) |
| Pull interface | Ø 7 mm |
| Gas interface | Ø 8 mm |
| Power supply for cooler and heating tube | power line |
| Operating time | 24 h |
| Data Storage | 100 sets of data |
| communication interface | USB , Infrared, optional: Bluetooth |
| Certification standards |
BlmSchV, KÜO (TÜV ByRgG 247), DIN EN50379-2 |
| defects liability period | 12 months |