Eye of Adsorption - Material Cycle Life Evaluation Analyzer
BSD-660 fully automatic high-throughput high-performance specific surface area and pore size analyzer
Product Overview:
Analysis of specific surface area and mesopores, micropores, and ultramicropores
High throughput analysis: up to 12 samples can be analyzed simultaneously
◆ Fully automatic: degassing → testing, fully automatic switching, no need for human intervention
Main Function
High throughput rapid specific surface area analysis;
◆ Pore volume and pore size distribution (mesoporous, microporous, ultramicroporous)
Conventional gas adsorption, such as N2, O2, Ar, CO, CO2, etc
◆ Adsorption of combustible gases, such as H2, CH4, C2H6 and other alkynes;
Technical Parameters
Wide testing range: specific surface area above 0.0005 ㎡/g, pore size 0.35-500nm;
High testing accuracy: specific surface area, pore size, pore volume, adsorption capacity ≤ 1% RSD (standard sample);
◆ Program heating and degassing: Software controlled program heating, room temperature -400 ℃, accuracy better than 0.1 ℃;
Intelligent degassing completion judgment: supports software automatic judgment, automatically judges the degassing effect based on pressure changes;
◆ Anti flying degassing: "Program controlled pressure"+"Program controlled temperature"+"Degassing furnace lifting"="Pressure controlled temperature rise";
Vacuum degree: 10-2Pa, optional molecular pump, vacuum degree can reach 10-8Pa;
Eye of Adsorption - Material Cycle Life Evaluation Analyzer
Technical Advantages
◆ High throughput: Supports analysis of up to 12 samples at a time;
◆ Fully automated: The degassing furnace and Dewar flask switch automatically, without the need for manual transfer of sample tubes or degassing furnaces;
High time utilization rate: It solves the problem of wasting time when conventional instruments cannot start testing after degassing is completed after work, making it a reality to install samples after work and check data at work;
Eliminate "helium pollution": helium gas testing dead volume → vacuum heating degassing → adsorption testing solves the problem of helium pollution in microporous analysis internationally and improves testing accuracy;
◆ Anti flying degassing: Supports "programmed pressure control" and "programmed temperature control" degassing, automatically raising and lowering the degassing furnace according to pressure changes to prevent sample flying;
◆ Support automatic cycle testing: automatic degassing+test cycle testing, used to evaluate the stability and lifespan of material adsorption performance;
◆ "Pressure controlled heating" anti flying degassing technology;
◆ "Automatic constant temperature zone" technology;
◆ Convenient installation and sealing: The one-time sealing technology for 6 sample tubes in a single analysis station eliminates the need to seal each tube individually.
◆ Full temperature control of the pneumatic system: The internal pneumatic system of the instrument is fully temperature controlled to 40 ℃, with an accuracy better than 0.1 ℃;
◆ Upward sliding door: Humanized and easy to open and close, saving laboratory space;
◆ Electric turbine liquid nitrogen pump: Humanized liquid nitrogen addition, stepless speed regulation, free movement, safe and convenient, liquid nitrogen pollution-free;
◆ High reliability: International supplier system, all core components are adopted;