- Phone
-
Address
No. 56-39 Huanghe North Street, Huanggu District, Shenyang City, Liaoning Province
Shenyang Jiaye Scientific Instrument Co., Ltd
No. 56-39 Huanghe North Street, Huanggu District, Shenyang City, Liaoning Province
Electronic industry cleanliness level testing particle counter3920
High flow sampling (50L/min), efficient and fast
Simultaneous testing of 6 particle sizes (0.5, 0.7, 1.0, 3.0, 5.0, 10.0 μ m)
External sensors can test environmental temperature, relative temperature, wind speed, and differential pressure (version only)
6.4-inch true color touch screen
Can store 10000 sets of data, and can also be connected to an external USB flash drive to increase storage capacity
Built in printer, directly output test report
Directly determine whether the cleanroom level meets the requirements
Compliant with JJF1190-2008 standard, the counting report complies with GB/T16292-2010 and ISO14644-1 standards
Electronic industry cleanliness level testing particle counter3920
name call, name, say, weigh |
Kanomax dust particle counter |
||
type number |
3905 |
3910 |
3920 |
flow quantity |
28.3L/min |
50L/min |
100L/min |
Test particle size |
Standard:0.3 、0.5、1.0、3.0、5.0The10.0μm Non standard:0.3~10.0μmChoose from within6Seed particle size |
Standard:0.5 、0.7、 1.0、3.0、5.0The10.0μm Non standard:0.5~10.0μmChoose from within6Seed particle size |
|
Counting efficiency |
50±20%@0.3μm conform toISO21501-4 |
50±20%@0.5μm conform toISO21501-4 |
|
Maximum particle concentration |
500,000a/cf |
300,000a/cf |
|
particle sizeresolution |
<15%@0.3μm conform toISO21501-4 |
<15%@0.5μm conform toISO21501-4 |
|
Test standard |
ISO14644-1TheISO14644-2015、new versionGMPTheGB/T16292-2010TheGB50333-2013TheFEDERAL STANDARDTheBRITISH STANDARDTheEC GMP |
||
language words |
Chinese-English translation |
||
light source |
laser diode |
||
Consistency loss |
<150000a/cf 5% >150000a/cf 10%conform toISO21501-4 |
||
Miscounting |
0.2a/cfThe following are in complianceISO21501-4 |
||
zero plan number |
conform toJISstandard |
||
apparent show |
6.4inchTrue color touch screen |
||
test method |
single timeIntervalcontinuous Standards and statistics |
||
data security |
conform to21 CFR Part11 |
||
data storage |
10,000Group data (number of files), can also be externally connectedUCopy the disk |
||
Map coding |
On site drawings0~999a |
||
errorapparentshow |
Concentration overflow, abnormal light source, abnormal flow rateExceeding the limit value |
||
audio output |
Built in speaker, capable of providing both alarm and error audio outputs |
||
pick up 口 |
USBTheEthernet |
||
hit India |
built-in printer |
||
sampling time |
6sec~23h59min 59sec |
||
sampling times |
repeated test1~9,999time |
||
Counting alarm |
in1~9,999,999aWithin the counting range |
||
External sensors |
Wind speed, temperature and humidity, pressure |
||
electricity pool |
Rechargeable lithium battery (two battery slots) |
||
electricity source |
100~240V AC50~60Hz |
||
quality guarantee period |
Two years for the whole machine, three years for the laser light source |
||
heavy quantity |
about4.5Kg(Excluding battery) |
about5.0Kg(Excluding battery) |
|
Overall dimensions |
200×205×200mm |
200×205×250mm |
|
Standard Appendix |
Operation manualTest reportZero count filterlithiumbattery×1、Waiting for the power sampling headClean pipe×2mTheACAdapterCommunication software, printing paper |
||
can select document |
environmentsensor(Wind speed, temperature and humidity, differential pressure)High pressure diffuser 、Scanning probe, portablebox |
||
Electronic industry cleanliness level testing particle counter3920