The AKD-A3 basic Karl Fischer micro moisture analyzer from Aikorod brand integrates multiple advanced technologies, and its measurement principle is the world-renowned classic method for measuring moisture, the Karl Fischer Coulomb method. It has a user-friendly interface, a 128 * 64 blue screen LCD display, and all Chinese dishes are available

Product Details





Acored brandThe AKD-A3 basic Karl Fischer micro moisture analyzer integrates multiple advanced technologies and uses the world-renowned classic method for measuring moisture, the Karl Fischer Coulomb method. It has a user-friendly interface, a 128 * 64 blue screen LCD display, and a fully Chinese menu. It has the advantages of high measurement accuracy, good repeatability, and low testing cost. The detection circuit adopts digital processing technology, which overcomes the interference of temperature drift and power supply voltage fluctuations in traditional instruments on the testing process, and improves the accuracy and sensitivity of the testing. The whole machine adopts a relay free process design, eliminating the hidden dangers caused by the oxidation of memory relay contacts and extending the service life of the instrument. It has been widely used in sectors such as power, petroleum, chemical, and pharmaceutical, and is one of the cost-effective products launched by our company.



Acored brandAKD-A3 Basic Karl Fischer Micro Moisture AnalyzerTechnical parameters:

model: AKD-A3
Measurement method: Karl Fischer Coulomb coulometric method
According to the standard: GB7600-87
Display measurement range: 0.1ug water -200mg water
Water content range: 0.0001%(1ppm)—100%
resolving power: 0.1ug water
Display mode: LCD blue screen display
Accuracy: When the water content is between 3ug and 1000ug, the measurement error is ≤± 2ug
When the water content is above 1000ug of water, the measurement error is ≤± 0.2% (excluding injection error)
Endpoint indication: Screen display \ \ Sound warning \ \ End light prompt
Blank processing: The blank current microprocessor automatically controls compensation to ensure accurate deduction of blank during the 10 minute sample enrichment time
Data input: key-in
Electrolysis speed: ≤ 2.0 mg water/min
Electrolytic current: 0-400mAh
Calendar/Clock: Analyze time and date display (without losing power)
External dimensions: 280*300*220mm
Power consumption: 30W
Using power supply: Voltage 220V ± 10%, frequency 50HZ ± 2.5HZ
Usage environment: Temperature -5-40 ℃, humidity<85%
Weight: About 5KG


Acored brandAKD-A3 Basic Karl Fischer Micro Moisture AnalyzerScope of application:

liquid Pharmaceutical industry: raw materials for traditional Chinese and Western medicine, etc
Pesticide industry: emulsification, etc
Petroleum and power industry: insulation oil, transformer oil, etc
Chemical industry: acids, alcohols, benzenes, phenols, ethers, organic solvents, etc
Other industries: lithium batteries, electrolytes, etc
solid Solid substances with good solubility, such as various inorganic salts and citric acid
gas Natural gas, liquefied gas, fluorocarbons, butadiene and other gases


Acored brandAKD-A3 basic Karl Fischer micro moisture analyzer meets the standard:
GB/T 6283-2008 Determination of Moisture Content in Chemical Products (Karl Fischer Method)
ASTM E1064-2008 Karl Fischer Coulombic Titration for Determination of Water Content in Organic Liquids
GB/T 7600-1987 Determination of Moisture Content in Operating Transformer Oil (Coulombic Method)
ASTM D4928-00 (2010) Karl Fischer Coulombic Titration for Determination of Water Content in Crude Oil
ASTM D6304-2007 Karl Fischer coulometric titration method for determining water content in petroleum products, lubricants, and additives
ISO 10337-1997 Determination of Moisture in Crude Oil (Karl Fischer Coulombic Titration Method)
GB/T 11146-2009 Determination of Crude Oil Water Content (Karl Fischer Coulombic Titration Method)
GB/T 3727-2003 Determination of Trace Water in Ethylene and Propylene for Industrial Use
GB/T 5074-1985 Method for determination of moisture content in coking products by microcoulometry
GB/T 6023-2008 Determination of Trace Water in Industrial Butadiene (Karl Fischer Coulombic Method)
GB/T 7376-2008 Determination of Trace Moisture in Fluoroalkanes for Industrial Use (Karl Fischer Method)
GB/T 18619.1-2002 Determination of Water Content in Natural Gas (Karl Fischer Coulombic Method)
GB/T 18826-2002 Industrial 1,1,1,2-tetrafluoroethane HFC-134a
SH/T 0246-1992 Determination of Water Content in Light Petroleum Products (Electricity Method)
SH/T 0255-1992 Method for determination of moisture content in additives and lubricating oils containing additives (electric method)
ASTM E1064-2008 Karl Fischer Coulombic Titration for Determination of Water Content in Organic Liquids
ASTM D4928-00 (2010) Karl Fischer Coulombic Titration for Determination of Water Content in Crude Oil
ASTM D6304-2007 Standard for Determination of Water Content in Petroleum Products, Lubricants, and Additives by Karl Fischer Coulombic Titration


Acored brandAKD-A3 Basic Karl Fischer Micro Moisture AnalyzerHas the following characteristics:
1. 128 * 64 blue screen LCD display, fully Chinese interactive menu, real-time display of instrument working status;
2. Adopting a composite control system of microcontroller and computer, dual CPU design, intelligent analysis and determination of moisture content;
3. The dual power channel separates the electrolytic electrode from the measuring electrode, and the instrument can automatically control various interferences, greatly improving the accuracy of test results;
4. The blank current microprocessor automatically controls compensation, and the reagent can quickly reach equilibrium state;
5. The instrument has a self checking function. If there is a short circuit or open circuit fault in the electrode, the instrument will automatically prompt the user;
6. 0-400mAh high electrolysis current, high detection sensitivity and fast analysis speed;
7. Surface electrostatic spraying treatment, anti-corrosion, easy to clean;
8. The perpetual calendar function includes year, month, day, week, hour, minute, and second, which are automatically stored after power failure and will not be lost for 10 years.


Acored brandAKD-A3 Basic Karl Fischer Micro Moisture AnalyzerOperation steps:
1. After installing the instrument accessories according to the instructions, turn on the instrument;
2. Set parameters according to the instructions, select different setting items, and press the confirm button to save after selection;
3. Reagent equilibrium. Inject an appropriate amount of pure water into a 50 microliter sampler to neutralize any excess water in the reagent;
4. Instrument calibration. Extract 2ul of pure water using a 10ul sampler and inject it. The displayed result should be "2000 ± 200ug (excluding injection error)". If the displayed result is within the error range, sample measurement can be performed.
5. Sample determination. After pressing the start button to inject the sample, the instrument will automatically perform the test. After the test is completed, a beep alarm will sound and the display screen will indicate "Test completed".


Acored brandAKD-A3 Basic Karl Fischer Micro Moisture Analyzermatters needing attention:
1、 Cannot display normally
Please check if the power connection wire, fuse, and power switch of the instrument are functioning properly.
2、 The instrument displays iodine
1. Evaluate whether the reagent is over iodine. If so, use a 0.5 microliter sampler to extract 0.2-0.4 microliters of water and inject it. Do not use an injector of 100 microliters or larger for injection.
2. Check whether the platinum wire at the bottom of the measuring electrode is connected together, causing a short circuit.
3、 The instrument shows an open circuit
Check if the measuring electrode is in good contact and re insert it firmly. Check if the connecting wire is broken.
4、 Open electrolysis without counting
Check if the electrolytic electrode is in good contact, unplug and plug it again, and check if the connecting wire is broken.
5、 Electrolysis does not end
Check if the reagent has expired.