| Measurement project | |
| Measurement project | Piston outer circle: cross-sectional profile, longitudinal profile, coaxiality Piston pin hole: unilateral pin hole roundness, cylindricity, longitudinal section line measurement Piston ring angle Roundness, cylindricity, diameter, taper (vertical, plane), concentricity, coaxiality (multi-element), straightness, perpendicularity (multi-element), Bounce (axial, radial), full bounce (axial, radial), flatness (multi-element), parallelism (multi-element), notch removal, waviness, Spectral analysis, wave height analysis, harmonic analysis, wall thickness deviation |
| Spindle Precision | |
| Radial error of spindle H=to the height of the workbench |
±(0.025+3H/ 10000)μm |
| Axial error of spindle X=distance to the center of the spindle |
±(0.025+6X/10000)μm |
| workbenches | |
| Table diameter | 220mm |
| adjustment range | Adjust the eccentricity by ± 2mm; Adjust the level ± 1 ° |
| Rotation diameter | 400mm |
| Capacity | 40Kg |
| rotational speed | 0-12 revolutions per minute |
| ZAxis (Column) | |
| Straightness (narrow range) | 0.35μm /100mm |
| Straightness (full range) | 0.6μm /400mm |
| Positioning speed | 1mm/s-8mm/s |
| Grating resolution | 1μm |
| Rotation axis and ZParallelism of axis guide rail | |
| Rotation axis and Z-axis guide rail parallelism |
2 μ m/400mm (busbar reference) |
| Horizontal arm (horizontal guide rail) | |
| Mobile accuracy | 1.5μm/150mm |
| Horizontal movement distance | 180mm |
| Positioning speed | 1mm/s-5mm/s |
| Grating resolution | 1μm |
| sensor | |
| range | 500 μ m (radius difference) |
| resolution | 0.005μm |
| Measuring needle shape | 2mm gemstone ball measuring head (with more diameters to choose from) |
| dynamometry | 1-12g |
| measuring range | |
| Maximum measurement diameter | 500mm |
| Maximum measurement height | 400mm |
| maximum measuring depth | Using standard probe: 100mm (when aperture is less than 36mm) Non standard bracket can measure up to 300mm (aperture greater than 36mm (optional)) |
| data acquisition | |
| data acquisition | 7200 points per week (internationally recognized) |
| Magnification factor | |
| Magnification factor | 200000 times automatic |
| appraising method | |
| appraising method | LSC (least squares method), MZC (minimum area method), MIC (maximum inscribed circle method), MCC (minimum circumcircle method) |
| Filter value | |
| Filter value | 1upr,1-15upr,1-50upr,1-150upr,1-500upr |
| Data analysis and processing | |
| Spectrum analysis | Analyze the amplitude of different frequency components |
| Abnormal data removal | Manually or automatically remove abnormal data such as burrs, holes, and bumps |
| file management | Measurement data is automatically saved and can be deleted |
| Data printing | Single or multiple measurement results can be printed on the same page in the same printing mode |
| INSTALLATION REQUIREMENTS | |
| power supply | 220-240V AC, 50HZ, Separate ground wire; Maximum power consumption 450W |
| compressed air | Pure compressed air: pressure 0.4MP, consumption 0.2m3/min (air supply 0.6MP, 0.25m3/min or above) |
| Environmental requirements | Working temperature: 10 ° -30 ° C (normal temperature); Humidity 20-80% RH (no condensation), good grounding, no obvious vibration source |