E+H temperature transmitter

NegotiableUpdate on 05/24
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

E+H Temperature Transmitter TMT181-AAAAA: In stock supply of original German E+H transmitters. The set type transmitter has high reliability, accuracy, and long-term stability, and can be used in key processes in all industries. It can not only transmit signals converted from resistance temperature detectors (RTDs) and thermocouples (TCs), but also transmit resistance and voltage signals.

Product Details

E+H temperature transmitterIntroduction to TMT181-AAAAA:

The German original E+H transmitter TMT181-AAAAA is available in stock. The set type transmitter has high reliability, accuracy, and long-term stability, and can be used in key processes in all industries. It can not only transmit signals converted from resistance temperature detectors (RTDs) and thermocouples (TCs), but also transmit resistance and voltage signals.

High precision measurement across the entire temperature range of the environment.

When the sensor is open or short circuited, it will emit a fault signal, pre-set to comply with NAMUR NE 43 standard.

Explosion proof certification: ATEX Ex ia and Zone 22 dust explosion-proof certification in accordance with EN 50281-1, FM IS, and CSA IS standards.

EN 50281-1; FM IS; CSA IS.

Compliant with NAMUR NE 21 standard for electromagnetic compatibility (EMC) and CE certification.

Use setting tools for online configuration during the measurement process.

Electrical isolation.

analog outputE+H temperature transmitterTMT181-AAAAA.

Temperature Transmitter.

TMT181-AAAAA Measurement Principle Head Transmitter.

Input RTD, TC, Ohm, mV.

Output 4... 20 mA.

Power supply 8... 35 V DC (standard type), 8... 30 V DC (explosion-proof type).

Communication PCP (PC Programmable).

Install B-type terminal.

Measurement accuracy (Pt 100, -50... 200 ℃)<=0.2 K.

Electrical isolation is.

ATEX certification: (Ex ia); FM+CSA: (IS); GL classification certification; UL complies with the 3111-1 standard.